In Vitro Fit Accuracy of Conical Connection Implant System Components - Part 1: Two-Dimensional Tolerance Displacement.
Charles Beng Kai Tan, Ming Yi Tan, Yoong Liang See Toh, Janice Zhi Hui Tan, Keson Beng Choon Tan
Key points
From the paper's conclusion- Statistically significant differences were observed among implant systems. Mean dr of all conical connection implant system components evaluated ranged from 2.7 to 23.7 μm. These measured displacements are interpreted to be small in magnitude, but their clinical significance remains uncertain.
Abstract
PURPOSE: This in vitro study evaluated the two-dimensional (2D) tolerance displacement of six conical connection implant systems and investigated the effect of varying tightening torque levels at four implant component connections. MATERIALS AND METHODS: Six implant systems were tested using ten implants, ten replicas, ten impression copings and ten abutments at four component connections: implant-impression coping (I-C), replica-impression coping (R-C), replica-abutment (R-A), and implant-abutment (I-A). A coordinate measuring machine was used to compute 2D tolerance displacement (dr) between upper (C / A) and lower (I / R) components, representing 2D positional deviation. Torque magnitudes of 5, 10, and 15 Ncm were applied at I-C, R-C, and R-A, while manufacturer-specified torque values were used for I-A. For I-C, R-C and R-A, the data was analyzed using mixed analysis of variance (ANOVA) to evaluate the effects of implant system and torque magnitude on dr. For I-A, one-way ANOVA was conducted to evaluate differences in dr among the specific system-abutment combinations at their respective manufacturer-specified torque values. RESULTS: Overall mean dr ranged from 2.7 ± 1.3 μm (CONR-A, 10 Ncm) to 23.7 ± 10.4 μm (MEGR-A, 10 Ncm). Mean dr ranged from 4.0 ± 2.7 μm (DENI-C, 10 Ncm) to 15.2 ± 8.8 μm (STRI-C, 10 Ncm) for I-C; from 3.9 ± 1.9 μm (DENR-C, 5 Ncm) to 13.0 ± 11.3 μm (STRR-C, 10 Ncm) for R-C; and from 2.7 ± 1.3 μm (CONR-A, 10 Ncm) to 23.7 ± 10.4 μm (MEGR-A, 10 Ncm) for R-A. Mixed ANOVA demonstrated no significant effects of torque magnitude or torque and implant system interaction for I-C, R-C, or R-A, but significant differences among implant systems were observed for all three connection types. For I-C, STR exhibited greater dr than ANT and DEN; for R-C, MEG exhibited greater dr than DEN and OSS; and for R-A, CON exhibited lower dr than all other systems. Significant differences among systems were also observed for I-A (P < 0.001). CONCLUSIONS: Statistically significant differences were observed among implant systems. Mean dr of all conical connection implant system components evaluated ranged from 2.7 to 23.7 μm. These measured displacements are interpreted to be small in magnitude, but their clinical significance remains uncertain.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In Vitro Fit Accuracy of Conical Connection Implant System Components - Part 2: Vertical Fit.The International journal of oral & maxillofacial implants · 2026
- A comparative analysis of efficacy of torque limiting devices provided by different dental implant manufacturersBMC Oral Health · 2026
- Accuracy, insertion torque and ISQ of keyless versus drill-key guided implant surgery: in vitro comparison of two implant macro-designs in healed ridges and molar extraction sites.Journal of dentistry · 2026
- Effect of Progressive Versus Standard Implant Thread Designs on Primary Stability and Insertion Torque: An In Vitro StudyThe International Journal of Oral & Maxillofacial Implants · 2026
- Influence of 3D Printing, Implant Macro-Design and Alveolar Ridge Morphology on Accuracy and Primary Stability in Static Computer-Assisted Implant Surgery: An In Vitro Study.Journal of dentistry · 2026
- Linear and Angular Deviations of Complete-Arch Digital Implant Impressions Using Three Scan Body Systems: An In Vitro Study.Dentistry journal · 2026