Influence of 3D Printing, Implant Macro-Design and Alveolar Ridge Morphology on Accuracy and Primary Stability in Static Computer-Assisted Implant Surgery: An In Vitro Study.
Malte Bagratuni, Samir Abou-Ayash, Justus Pook, Stefan Wentaschek, Lauren Bohner, Gülce Cakmak, Florian Beuer, Stefano Pieralli
Department of Prosthodontics, Geriatric Dentistry and Craniomandibular Disorders, Charité - University Medicine Berlin, Berlin, Germany. Electronic address: malte.bagratuni@charite.de. · Department of Prosthetic Dentistry and Material Science, University Medical Center of the Johannes Gutenberg University Mainz, Mainz, Germany.
Key points
From the paper's conclusion- In vitro, DLP and ME guides showed no significant difference in positional accuracy. TLC achieved higher IT and ISQ but greater deviation than TLX. Deviations were smaller at healed premolar ridges than at molar extraction sites.
Abstract
OBJECTIVES: To assess the positional accuracy of static computer-assisted implant surgery (sCAIS) using guides printed by digital light processing (DLP) or material extrusion (ME), and to compare insertion torque (IT) and implant stability quotient (ISQ) of two tissue-level implants (Straumann TLC, TLX) in healed premolar ridges versus molar extraction sites in vitro. METHODS: Eighty implants were placed in standardized resin bone models using DLP- or ME-printed guides. Angular, crestal, and apical deviations were determined by superimposing preoperative and postoperative scans; IT and ISQ were recorded. Statistical analysis was performed with three-way ANOVA, aligned rank transform ANOVA and ordinal logistic regression. RESULTS: Healed premolar ridges showed smaller angular (2.35 ± 1.36° vs. 3.33 ± 1.62°), crestal (0.50 ± 0.20 vs. 0.75 ± 0.36 mm), and apical deviations (0.86 ± 0.41 vs. 1.33 ± 0.63 mm) than molar extraction sites (p<0.01). Printing technology did not affect accuracy, but ME guides showed lower IT. TLC deviated more than TLX but reached higher ISQ (81.58 ± 6.03 vs. 77.18 ± 5.18; p<0.01) and IT. CONCLUSIONS: In vitro, DLP and ME guides showed no significant difference in positional accuracy. TLC achieved higher IT and ISQ but greater deviation than TLX. Deviations were smaller at healed premolar ridges than at molar extraction sites. CLINICAL SIGNIFICANCE: ME-printed guides showed accuracy comparable to DLP-printed guides in vitro, however clinical equivalence requires in vivo confirmation. Larger deviations should be anticipated at molar extraction sites, with adequate safety margins. Implant macro-design affected accuracy and primary stability in opposite directions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy, insertion torque and ISQ of keyless versus drill-key guided implant surgery: in vitro comparison of two implant macro-designs in healed ridges and molar extraction sites.Journal of dentistry · 2026
- Does manufacturing specialization matter? Transfer accuracy of static surgical guides produced by general-purpose and dental-specific additive workflows: A paired in vitro study.Journal of clinical and experimental dentistry · 2026
- Comparative Evaluation of SLA and DLP 3D Printing in Dental Implant Guides: Impact on Fabrication Accuracy, Speed, and Resin UsageDentistry Journal · 2025
- Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.Clinical oral implants research · 2026
- Accuracy of Computer-Guided Dental Implant Placement: A Clinical Comparison of Three Surgical Guide TypesJournal of Clinical Medicine · 2025
- Short implant placement using dynamic and robotic computer-assisted implant surgery (d-CAIS and r-CAIS): a controlled in vitro study.Journal of dentistry · 2026