In Vitro Fit Accuracy of Conical Connection Implant System Components - Part 2: Vertical Fit.
Charles Beng Kai Tan, Ming Yi Tan, Yoong Liang See Toh, Janice Zhi Hui Tan, Keson Beng Choon Tan
Key points
From the paper's conclusion- Vertical fit accuracy of conical connection implant components is influenced by implant system, component connection, and torque magnitude, with system-specific differences. No single tightening torque can consistently reproduce abutment nominal heights between replica-abutment and implant-abutment interfaces across all implant systems.
Abstract
PURPOSE: This in vitro study evaluated the vertical fit accuracy of six conical connection implant systems and investigated the effect of varying tightening torque levels at four implant component connections. MATERIALS AND METHODS: Six implant systems were tested at four component connections: implant-impression coping (I-C), replica-impression coping (R-C), replica- abutment (R-A), and implant-abutment (I-A). Ten implants (I), replicas (R), open-tray impression copings (C), and abutments (A) from each system were randomly selected for testing. Torque levels of 5, 10, and 15 Ncm were applied to I-C, R-C, and R-A connections, while manufacturer-specified torque values were used for I-A. Changes in upper component (C / A) nominal height (ΔNH for I-C, R-C, R-A; ΔNHA for I-A) were measured using a coordinate measuring machine and represented vertical fit. Data was analysed using mixed analysis of variance (ANOVA) to evaluate the effects of implant system and torque magnitude on vertical fit, which was quantified by ΔNH or ΔNHA (α = 0.05). RESULTS: Significant effects of torque magnitude, implant system, and their interaction on vertical displacement were observed for all component connections (P < 0.001). For I-C, R-C, and R-A, increasing torque from 5 to 10 and 15 Ncm generally resulted in decreased upper component nominal height. Mean ΔNH ranged from -0.3 ± 0.5 µm (CON, ΔNH10) to -13.9 ± 2.6 μm (ANT, ΔNH15) for I-C; from -0.3 ± 0.5 µm (CON, ΔNH10) to -25.9 ± 6.0 µm (ANT, ΔNH15) for R-C; and from -2.5 ± 1.6 µm (MEG, ΔNH10) to -33.9 ± 3.3 µm (CON, ΔNH15) for R-A. Mean ΔNHA ranged from -21.8 ± 9.7 μm (DEN, ΔNHA5) to 76.4 ± 25.4 μm (CON, ΔNHA15). ΔNHA was consistently positive for CON, MEG, and OSS, consistently negative for DEN and STR, and torque dependent for ANT. CONCLUSION: Vertical fit accuracy of conical connection implant components is influenced by implant system, component connection, and torque magnitude, with system-specific differences. No single tightening torque can consistently reproduce abutment nominal heights between replica-abutment and implant-abutment interfaces across all implant systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In Vitro Fit Accuracy of Conical Connection Implant System Components - Part 1: Two-Dimensional Tolerance Displacement.The International journal of oral & maxillofacial implants · 2026
- Effect of Progressive Versus Standard Implant Thread Designs on Primary Stability and Insertion Torque: An In Vitro StudyThe International Journal of Oral & Maxillofacial Implants · 2026
- Accuracy, insertion torque and ISQ of keyless versus drill-key guided implant surgery: in vitro comparison of two implant macro-designs in healed ridges and molar extraction sites.Journal of dentistry · 2026
- Interaction Between Implant Macrogeometry and Osteotomy Preparation on Immediate Mechanical Fixation and Regional Thread Alterations: An Ex Vivo Biomechanical Study.Bioengineering (Basel, Switzerland) · 2026
- Marginal Bone Loss and Peri-Implant Clinical Outcomes of Internal Hexagonal and Internal Conical Implant Abutment Connections: A Narrative Review.Cureus · 2026
- A comparative analysis of efficacy of torque limiting devices provided by different dental implant manufacturersBMC Oral Health · 2026