Accuracy, insertion torque and ISQ of keyless versus drill-key guided implant surgery: in vitro comparison of two implant macro-designs in healed ridges and molar extraction sites.

Malte Bagratuni, Gülce Cakmak, Omar Elattar, Samir Abou-Ayash, Stefan Wentaschek, Florian Beuer, Lauren Bohner, Stefano Pieralli

Department of Prosthodontics, Geriatric Dentistry and Craniomandibular Disorders, Charité - Universitätsmedizin Berlin, Aßmannshauser Straße 4-6, 14197, Berlin, Germany. · Department of Prosthodontics, Geriatric Dentistry and Craniomandibular Disorders, Charité - Universitätsmedizin Berlin, Aßmannshauser Straße 4-6, 14197, Berlin, Germany; Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland; Faculty of Dentistry, Biruni University, Istanbul, Turkey. Electronic address: guelce.cakmak@charite.de.

Preclinical

Key points

From the paper's conclusion
  • The keyless system yielded lower linear deviations and higher precision than the drill-key system, whereas both macro-designs achieved comparable trueness. TLC reached higher IT and ISQ than TLX. Healed ridges provided greater accuracy and primary stability than simulated molar extraction sites.

Abstract

OBJECTIVES: To compare positional accuracy, insertion torque (IT) and implant stability quotient (ISQ) between a keyless and a drill-key system for static computer-assisted implant surgery (sCAIS), with two implant macro-designs in healed ridges (HR) and molar extraction sites (ES). METHODS: Eighty tissue-level implants (Straumann) of two macro-designs (TLC, TLX) were placed in standardized maxillary posterior bone models with HR and ES using a keyless or a drill-key sCAIS system. Angular, crestal and apical deviations were evaluated for trueness and precision by superimposing pre- and postoperative scans. IT and ISQ were recorded and analyzed using three-way and ART ANOVA and ordinal logistic regression (α=0.05). RESULTS: The keyless system produced significantly lower crestal (p=0.02) and vertical deviations at the shoulder (p<0.01) and apex (p=0.01) than the drill-key system, with higher precision. Implants in HR showed smaller angular, crestal and apical deviations than those in ES (all p<0.01). IT was higher with the drill-key system, with TLC and in HR (all p≤0.01). CONCLUSIONS: The keyless system yielded lower linear deviations and higher precision than the drill-key system, whereas both macro-designs achieved comparable trueness. TLC reached higher IT and ISQ than TLX. Healed ridges provided greater accuracy and primary stability than simulated molar extraction sites. CLINICAL SIGNIFICANCE: Under in vitro conditions, a keyless sCAIS system transferred the planned implant position with smaller crestal and vertical deviations than a drill-key system. In simulated molar extraction sites, larger positional deviations may be anticipated with both implant macro-designs. Primary stability data obtained in vitro, in particular ISQ values, cannot be extrapolated to the clinical situation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers