Feasibility of Ultra-Low-Dose Photon-Counting CT for Anatomy-Based Cochlear Implant Fitting: A Cadaveric Proof-of-Concept Study.
Thore Schade-Mann, Hubert Löwenheim, Tobias Albrecht
Department of Otolaryngology-Head & Neck Surgery, University of Tübingen Medical Center, Tübingen, Germany.
Key points
From the paper's conclusion- In this single-donor proof-of-concept study, PCD-CT at 5% of the institutional reference dose retained electrode visibility and yielded small frequency-estimation deviations for the evaluated intracochlear contacts. These findings therefore support the technical feasibility of substantial postoperative dose reduction within the evaluated electrode region. Validation using complete insertions in larger clinical cohorts is required.
Abstract
HYPOTHESIS: Stepwise reduction of CT tube current during postoperative photon-counting detector CT (PCD-CT) imaging preserves the frequency-estimation accuracy required for anatomy-based cochlear-implant fitting at doses as low as 5% of the clinical standard. BACKGROUND: Anatomy-based fitting (ABF) uses postoperative CT to allocate electrode contacts to the cochlear frequency map, reducing frequency-to-place mismatch and potentially improving cochlear implant (CI) outcomes. While PCD-CT lowers radiation compared with conventional CT, the minimum postoperative dose compatible with ABF has not been established. METHODS: A human body donor was imaged with PCD-CT (NAEOTOM Alpha, Siemens Healthineers) before and after CI test electrode insertion via a round window approach at 100%, 50%, 25%, 10%, and 5% tube current. Electrode angular insertion depth and frequency position were extracted using OTOPLAN software (Version 4, Cascination). Frequency deviations from the 100% reference were quantified in semitones. Analysis was repeated using the 50% preoperative scan as an alternative reference standard, extending previously published preoperative dose-reduction findings. RESULTS: Electrode detection was successful at all dose levels for the evaluated intracochlear contacts. Using the 100% preoperative reference, mean frequency deviation at 5% postoperative dose was 0.02±0.29 semitones, with no evaluated contact exceeding one semitone. When the 50% preoperative scan served as reference, deviations remained ≤1 semitone for postoperative doses ≥10%; three adjacent contacts marginally exceeded this threshold at 5% postoperative dose (maximum: -1.20 semitones). CONCLUSIONS: In this single-donor proof-of-concept study, PCD-CT at 5% of the institutional reference dose retained electrode visibility and yielded small frequency-estimation deviations for the evaluated intracochlear contacts. These findings therefore support the technical feasibility of substantial postoperative dose reduction within the evaluated electrode region. Validation using complete insertions in larger clinical cohorts is required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.Investigative radiology · 2026
- Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.European journal of radiology · 2026
- Feasibility of Electrode-Location Estimates Based on Preoperative CT-Imaging for Anatomy-Based Fitting in Cochlear Implant Recipients.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads.AJNR. American journal of neuroradiology · 2026
- Improved visualization of malpositioned middle ear prostheses using photon-counting detector CT: a cadaver study.Radiology advances · 2026