Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads.
Seong-Won Kim, Georgia R R S S Silva, Willem G Calderon, Ji Young Lee, Thad Benefield, Adam J Kimple, Yueh Z Lee, Kassie L McCullagh
From the University of North Carolina School of Medicine (S.-W.K.), Radiology (G.R.R.S.S.S., W.G.C., J.Y.L., T.B., Y.Z.L., K.L.M.), Otolaryngology- Head and Neck Surgery (A.J.K.), University of North Carolina, Chapel Hill, North Carolina, USA. · From the University of North Carolina School of Medicine (S.-W.K.), Radiology (G.R.R.S.S.S., W.G.C., J.Y.L., T.B., Y.Z.L., K.L.M.), Otolaryngology- Head and Neck Surgery (A.J.K.), University of North Carolina, Chapel Hill, North Carolina, USA. kassie_mccullagh@med.unc.edu.
Key points
From the paper's conclusion- PCCT maintains clinically sufficient temporal bone image quality at tube current-time products as low as 75 mAs, while doses of 25 mAs and below fall below the threshold for reliable clinical use. Per ALARA, this threshold optimizes safety via significant dose reduction while maintaining the high image quality necessary for complex temporal bone anatomy.
Abstract
BACKGROUND AND PURPOSE: Photon-counting computed tomography (PCCT) is an emerging technology that offers improved spatial resolution and noise reduction at lower radiation doses than conventional CT; however, the effect of dose reduction on clinically significant image quality has not been well established. Following the ALARA (As Low As Reasonably Achievable) principle, this study evaluated temporal bone image quality across decreasing radiation doses to identify the minimum dose required to maintain diagnostic integrity. MATERIALS AND METHODS: Seven cadaveric heads underwent temporal bone PCCT imaging at four tube current-time products: 290, 75, 25, and 11 mAs. All scans were acquired at 120 kV, 0.55 pitch, and a 0.5 second rotation time and reconstructed with 0.2-mm slice thickness in axial and coronal planes for left and right temporal bones. Three board-certified radiologists independently assessed prespecified, randomized temporal bone landmarks using a 5-point Likert scale to determine clinical utility (1-nondiagnostic; 5-excellent). Mixed-effects linear regression with random intercepts for patient, reader, and landmark estimated mean scores by dose. Pairwise comparisons were Tukey-adjusted, and one-sided FDR-adjusted t-tests evaluated whether predicted mean scores exceeded 3 (clinically sufficient). RESULTS: At 290 and 75 mAs, predicted mean scores were 4.81 (95% CI: 4.42-5.21) and 4.21 (95% CI: 3.82-4.61), respectively, demonstrating good/excellent image quality for all evaluated landmarks and exceeding threshold for clinical sufficiency (FDR-adjusted p<0.0001). At 25 mAs, predicted mean scores were 2.80 (95% CI: 2.41-3.19), falling in the poor/sufficient range, and at 11 mAs, were rated the lowest at 2.09 (95% CI: 1.70-2.48), scoring below the clinical sufficiency threshold (p=0.99). Overall, mean Likert image quality scores decreased with progressive mAs reduction, remaining technically interpretable but below the threshold for clinical sufficiency at 25 mAs. All pairwise comparisons were statistically significant (Tukey-adjusted p<0.0001). Interreader agreement (Krippendorff's α) ranged from 0.301 to 0.657 across landmarks. CONCLUSION: PCCT maintains clinically sufficient temporal bone image quality at tube current-time products as low as 75 mAs, while doses of 25 mAs and below fall below the threshold for reliable clinical use. Per ALARA, this threshold optimizes safety via significant dose reduction while maintaining the high image quality necessary for complex temporal bone anatomy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.European journal of radiology · 2026
- Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.Investigative radiology · 2026
- Optimization of the reconstruction kernel for temporal bone imaging using photon-counting detector CT: A combined physical and visual evaluationRadiography · 2025
- Optimization of reconstruction strategies for head and neck computed tomography angiography using photon-counting computed tomography: A comparative study with energy-integrating detector computed tomography.European journal of radiology · 2026
- Feasibility of Ultra-Low-Dose Photon-Counting CT for Anatomy-Based Cochlear Implant Fitting: A Cadaveric Proof-of-Concept Study.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026