Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.
Martin Georg Zeilinger, Stefan Lang, Markus Kopp, Jessica Graef, Maximilian Hinsen, Marco Wiesmueller, Lava Taha, Ulrich Hoppe,
Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany (M.G.Z., M.K., J.G., M.H., M.W., M.U., C.S.M.); Institute of Neuroradiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany (S.L., A.D.); Department of Otolaryngology-Head and Neck Surgery, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany (L.T., U.H., J.H., S.K.M.).
Key points
From the paper's conclusion- Regarding image quality in postoperative assessment of CI, PCCT is noninferior to or in single features better than FDCT at a significantly lower radiation dose and shorter scan time.
Abstract
OBJECTIVES: After FDA approval in 2021, photon-counting detector computed tomography (PCCT) has demonstrated its feasibility for high-resolution imaging while lowering the radiation dose at the same time. High-resolution visualization of cochlear implant (CI) integrity and positioning is crucial for postoperative management. The aim of this study was to evaluate the performance of PCCT for postoperative imaging after CI implantation in comparison to high-resolution flat-detector CT (FDCT), the current clinical gold standard. MATERIALS AND METHODS: Postoperative PCCT scans after cochlear implantation were retrospectively analyzed from August 2024 to April 2026. Scans were included for analysis if FDCT acquisition had been before PCCT on the same patient in the course of sequential bilateral cochlear implantation. Image quality was assessed in consensus by 2 experienced readers using a scoring system analyzing structural delineation of the inner and middle ear, CI integrity, metal and motion artifacts as well as calculating an overall score. Radiation dose was retrieved from the picture archiving and communication system. RESULTS: Twenty-eight postoperative PCCT scans of patients after CI implantation with a mean time delay of 4.5±5.1 years between FDCT and PCCT scans were evaluated. Image quality of PCCT was noninferior to FDCT for all single imaging features (P<0.001, eg, distinction of single CI electrodes: median [interquartile range] PCCT 3 [3 to 3]; FDCT 3 [3 to 3]; P<0.001) with the exception of the homogeneity of bony structures and the mastoid cell septation (P=0.999). PCCT was superior to FDCT in reducing metal and motion artifacts and increasing the delineation of the stapes (P=0.009 to <0.001; eg, metal artifacts PCCT 3 [2.75 to 3]; FDCT 2 [1 to 2]; P<0.001). Effective radiation dose of PCCT was 0.05±0.01 mSv and 8.8±5.4-times lower than for FDCT with 0.44±0.29 mSv (P<0.001) applying our institutional protocols. CONCLUSIONS: Regarding image quality in postoperative assessment of CI, PCCT is noninferior to or in single features better than FDCT at a significantly lower radiation dose and shorter scan time.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.European journal of radiology · 2026
- Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Feasibility of Ultra-Low-Dose Photon-Counting CT for Anatomy-Based Cochlear Implant Fitting: A Cadaveric Proof-of-Concept Study.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Improved visualization of malpositioned middle ear prostheses using photon-counting detector CT: a cadaver study.Radiology advances · 2026
- Clinical Photon-Counting CT in Neuroradiology: A Real-World Paired Comparison with Conventional Energy-Integrating CT.AJNR. American journal of neuroradiology · 2026
- Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart DiseaseDiagnostics · 2026