Improved assessment of the cochlear implant and temporal bone structures with ultra-high resolution photon-counting detector CT compared to conventional CT systems.
Judith van der Bie, Anke W van der Eerden, Aaike S van den Berg, Marcel L Dijkshoorn, Marcel van Straten, Hylke F E van der Toom, Floris V W J van Zijl, Joel de Groen,
Department of Radiology & Nuclear Medicine, Erasmus MC, Dr. Molewaterplein 40, 3015 GD, Rotterdam, The Netherlands. · Department of Radiology & Nuclear Medicine, Erasmus MC, Dr. Molewaterplein 40, 3015 GD, Rotterdam, The Netherlands. A.vandereerden@erasmusmc.nl.
Key points
From the paper's conclusion- Purpose This study evaluates the image quality of photon-counting detector CT (PCD-CT) compared to conventional CT systems in cochlear implant imaging and aims to determine optimal scan parameters.
Abstract
PURPOSE: This study evaluates the image quality of photon-counting detector CT (PCD-CT) compared to conventional CT systems in cochlear implant imaging and aims to determine optimal scan parameters. METHODS: Three cochlear implant models were inserted into three post-mortem human specimen heads. Each head was scanned on three high-end CT systems: SOMATOM Edge, SOMATOM Force and NAEOTOM Alpha. A total of 192 reconstructions were obtained, including various tube voltages (120-150 kV), tin filtering, and slice thicknesses (0.2-0.6 mm). For each scanner, multiple reconstruction kernels were applied. Contrast-to-noise ratios (CNR) for bone-to-soft-tissue were calculated. Two radiologists independently assessed anatomy, implant visibility, and overall image quality on a 5-point Likert scale and scored diagnostic confidence for detecting anatomic dehiscence. RESULTS: PCD-CT showed significantly higher image quality scores across implant, anatomy, and general image quality compared to the Edge and Force (p < 0.001). PCD-CT also produced significantly higher CNR (mean 23.5 vs. 13.8 [Edge] and 12.7 [Force]) and greater diagnostic confidence. For PCD-CT, optimal image quality was achieved using 0.2 mm slice thickness and the Hr89 kernel at 140 kV. PCD-CT scans acquired with half dose (27.5 mGy) showed no significant difference in CNRs but led to slightly lower subjective scores compared to standard-dose imaging. CONCLUSION: PCD-CT provides superior image quality and diagnostic confidence in cochlear implant imaging compared to the Edge and Force. Optimal results are achieved using 140 kV, 0.2 mm slices, and the Hr89 kernel. Although dose reduction is feasible, it slightly compromises subjective image quality.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.European journal of radiology · 2026
- Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.Investigative radiology · 2026
- Optimization of the reconstruction kernel for temporal bone imaging using photon-counting detector CT: A combined physical and visual evaluationRadiography · 2025
- Feasibility of Ultra-Low-Dose Photon-Counting CT for Anatomy-Based Cochlear Implant Fitting: A Cadaveric Proof-of-Concept Study.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads.AJNR. American journal of neuroradiology · 2026
- Improved visualization of malpositioned middle ear prostheses using photon-counting detector CT: a cadaver study.Radiology advances · 2026