Time-Efficient Visual Field Testing Using the Imo Visual Function Analyzer in Glaucoma.
Kenji Tanaka, Atsuya Miki, Robert N Weinreb, Yukiko Okabe, Fukuko Matsuoka, Shinya Eguchi, Masashi Nishigaki, Chiaki Nogawa,
Aichi Medical University Eye Center, Nagoya, Japan. · Hamilton Glaucoma Center, Shiley Eye Institute and Viterbi Family Department of Ophthalmology, University of California San Diego.
Key points
From the paper's conclusion- AIZE EX considerably reduced test time. Test results of AIZE EX strongly correlated with those of AIZE, but AIZE EX showed slightly higher sensitivity values. These results show AIZE EX effectively reduces test time without significantly compromising measurement accuracy, potentially reducing patient fatigue and improving acceptance of the test.
Abstract
PRCIS: Novel visual field measurement algorithm AIZE EX significantly reduced test duration without significantly compromising the accuracy. PURPOSE: To compare the accuracy and time of Ambient Interactive Zippy Estimation by Sequential Testing (AIZE) and time-saving Extra Mode (AIZE EX), using the imo visual function analyzer. PATIENTS AND METHODS: This retrospective study included 80 eyes from 59 patients with glaucoma who underwent visual field testing with both AIZE and AIZE EX strategies with the 24plus (1-2) test grid. Test duration, global indices (mean deviation [MD], pattern standard deviation [PSD], and visual field index [VFI]), and the number of significantly abnormal points on pattern deviation plots were compared. Differences in visual field sensitivity were analyzed at each individual test point. RESULTS: AIZE EX significantly reduced test duration by approximately 40% (202.6±35.2 vs. 358.5±76.3 seconds, P<0.001). Although AIZE EX showed significantly better MD (-6.9±7.3 vs. -7.7±7.2 dB), VFI (81.6±20.2 vs. 80.3±20.4%), and PSD (8.1±4.3 vs. 8.5±4.4 dB) (all P<0.001), differences were small. Strong correlations were observed between strategies for all global indices (P<0.001). Abnormal points on pattern deviation plots at P<2% and<5% were fewer with AIZE EXAIZE (P=0.005 and P<0.001). CONCLUSIONS: AIZE EX considerably reduced test time. Test results of AIZE EX strongly correlated with those of AIZE, but AIZE EX showed slightly higher sensitivity values. These results show AIZE EX effectively reduces test time without significantly compromising measurement accuracy, potentially reducing patient fatigue and improving acceptance of the test.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rapid, Non-Contact, Objective Multifocal Assessment of Visual Function in Glaucoma.Bioengineering (Basel, Switzerland) · 2026
- Predicting Variability and Reliability in Visual Field Testing: Short- and Long-Term ApproachesOphthalmology Science · 2026
- Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Feasibility and performance of eye movement perimetry for glaucoma triage in Tanzania: A prospective cross-sectional study in a real-world setting.PLOS global public health · 2026
- Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients.Frontiers in ophthalmology · 2026
- Accuracy and precision of visual acuity tests: A systematic review and meta-analysis across testing strategiesmedRxiv · 2026