Toward Indication-Stratified Local Reference Levels for IGRT Cone-Beam CT: A Clinical Audit of CTDIvol and DLP Across Ten Workflows on Versa HD.
O El Mouden, H Sekkat, A Bannan, F Chaara, H Bendouro, R El Baydaoui, M Bougteb, M Tahiri,
Laboratory of Sciences and Engineering of Biomedicals, Biophysics and Health, Higher Institute of Health Sciences, Hassan 1 University, Settat, Morocco. · University of Seville, Seville, Spain.
Key points
From the paper's conclusion- The reported indication-resolved /DLP distributions provide a practical basis for local reference levels and targeted optimization of high-dose workflows (pelvic/abdominal) while preserving low-dose protocols for cranial/extremity indications.
Abstract
To characterize routine kV cone-beam CT (CBCT) imaging dose in image-guided radiotherapy (IGRT) and to derive indication-specific local benchmark values using and DLP for clinical protocol optimization. This retrospective single-center audit included 300 CBCT examinations acquired on a Versa HD linear accelerator (Elekta AB, Stockholm, Sweden) using XVI v5.0.4. Examinations were grouped into 10 clinical indication workflows. For each CBCT, acquisition descriptors (CTDI phantom, kVp, total mAs, scan arc, FOV, bowtie, scan length) and dose indices were extracted. (mGy) was analyzed as recorded (reference value). DLP (mGy cm-1) was derived/verified as × scan length. Indication-stratified distributions were summarized using mean ± SD and 95% CI, with examination-level visualisation. An optional parameter-based estimation model () was evaluated as a quality-control layer using Bland-Altman analysis. Across all examinations, was 3.25 ± 3.06 mGy (range 0.123-10.109) with median 1.44 mGy (IQR 0.84-5.96). DLP was 80.27 ± 75.66 mGy cm-1 (range 2.21-255.81) with median 34.36 mGy cm-1 (IQR 18.55-150.38). Dose indices differed markedly by indication (≈49-fold between lowest and highest mean ). The highest mean occurred for Prostate IGRT (7.69 ± 1.95 mGy) and Pelvic nodes (7.21 ± 1.83 mGy), whereas Extremity setup (0.156 ± 0.027 mGy) and Brain SRS/SRT (0.263 ± 0.062 mGy) were the lowest. DLP was highest for Pelvic nodes (187.6 ± 47.5 mGy cm-1) and Prostate IGRT (169.1 ± 42.9 mGy cm-1). The estimation model showed minimal bias (Bland-Altman bias 0.007 mGy; 95% limits of agreement -0.321 to 0.335 mGy). Routine IGRT CBCT dose on Versa HD/XVI shows strong indication dependence, driven by protocol selection and scan extent. The reported indication-resolved /DLP distributions provide a practical basis for local reference levels and targeted optimization of high-dose workflows (pelvic/abdominal) while preserving low-dose protocols for cranial/extremity indications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Leveraging longitudinal ACR CT Dose Index Registry data to improve interpretation of institutional CT dose performance.Journal of applied clinical medical physics · 2026
- CT-ED/MD calibration curve commissioning and end-to-end validation of HyperSight CBCT-Based dose calculation for online adaptive radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Diagnostic Reference Levels for Dental Cone Beam Computed Tomography: A Critical Appraisal of Global Progress and Remaining Gaps.Cureus · 2026
- Technical benchmark of a high-speed ring gantry kV-CBCT imager for CBCT-based dose calculation.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Development of a pragmatic methodology for measurement of dose for KV CBCT in radiotherapy: A pilot study.Annals of the ICRP · 2026
- Chest computed tomography for oncological follow-Up: Clinical relevance and importance of dosimetric evaluation.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026