Technical benchmark of a high-speed ring gantry kV-CBCT imager for CBCT-based dose calculation.
Aurélien Badey, Hugo Le Brun, Enric Jaegle, Océane Bachelier, Véronique Bodez, Laure Vieillevigne, Catherine Khamphan
Department of Medical Physics, Sainte-Catherine Institut du Cancer-Avignon Provence, F-84000 Avignon, France. Electronic address: a.badey@isc84.org. · Department of Medical Physics, Sainte-Catherine Institut du Cancer-Avignon Provence, F-84000 Avignon, France.
Key points
From the paper's conclusion- These results confirm the value of 125 kVp HS images reconstructed with iCBCT Acuros CTS for CBCT-based dose calculation.
Abstract
PURPOSE: A CBCT platform (HyperSight (HS)) mounted on an O-ring linac (ETHOS v1.1), is compared with standard CBCT imagers installed on C-arm (TrueBeam v2.7) and O-ring (Halcyon v3.0) linacs, as well as planning CT (pCT). The study focuses on image quality, radiation dose, and dosimetric evaluation (Hounsfield unit (HU) accuracy and dose comparison). METHODS: Phantom-based comparisons were performed using default CBCT protocols and reconstruction algorithms available (FDK, iCBCT and iCBCT Acuros CTS). Quantitative image quality metrics (noise, CNR, TTF, low contrast resolution) were studied (IQMetrix software). Radiation dose measurements were compared between CBCT platforms and against manufacturer values.pCT-CBCT HU comparisons were performed on HU-density calibration and anthropomorphic phantoms (Mean Error (ME) and Mean Absolute Error (MAE)). Dosimetric evaluations included relative dose deviations (D50%), and 3D gamma analysis. RESULTS: The HS imager demonstrated improved image quality and detection efficiency associated with a reduction in radiation doses compared to standard CBCT imagers. HU accuracy improved markedly for HS images reconstructed with iCBCT Acuros CTS yielding MAE(Body) < 29HU except for Head, and |ME (Soft tissue)| <8HU except for Large Pelvis protocol. For density calibration phantom, HU deviations remained within ± 50 HU except for bone densities. Dosimetric comparisons showed dose deviations below 0.5% between pCT and CBCT-based calculations for all protocols using HyperSight with iCBCT Acuros CTS (except for Large Pelvis (0.9%)). Gamma passing rates were consistently above 97.7% for 0.5%/1 mm criteria. CONCLUSION: These results confirm the value of 125 kVp HS images reconstructed with iCBCT Acuros CTS for CBCT-based dose calculation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating Image Quality of Cone Beam Computed Tomography Using HyperSight® on the Varian TrueBeam.Annals of biomedical engineering · 2026
- HyperSight CBCT dose calculation accuracy for adaptive head and neck radiotherapy: Results from a prospective clinical trial.Medical physics · 2026
- Towards adaptive radiotherapy: Comparing HyperSight and standard CBCT reconstructions on a C-Arm Linac.Journal of applied clinical medical physics · 2026
- CT-ED/MD calibration curve commissioning and end-to-end validation of HyperSight CBCT-Based dose calculation for online adaptive radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Task-based image quality and dosimetric validation of HyperSight CBCT for adaptive radiotherapy.Journal of applied clinical medical physics · 2026
- Image quality robustness of HyperSight‐enabled versus conventional iterative cone‐beam CT across patient size and imaging geometryJournal of Applied Clinical Medical Physics · 2026