Development of a pragmatic methodology for measurement of dose for KV CBCT in radiotherapy: A pilot study.
M Djukelic, S Gros, T Kron, T Wood, C J Martin
Department of Medical Technology and Physics, Sir Charles Gairdner Hospital, Nedlands, WA, Australia; e-mail: mario.djukelic@genesiscare.com. · Department of Radiation Oncology, Cardinal Bernardin Cancer Center, Stritch School of Medicine, Loyola University Chicago, Maywood, IL, USA.
Key points
From the paper's conclusion- The results indicate that the alternative setup using equipment more accessible in radiotherapy facilities is a viable alternative to the use of a CT phantom to quantify kV CBCT radiation dose from linac-based image-guided radiotherapy equipment that can be used for optimisation processes.
Abstract
Imaging modalities such as on-board kilovoltage cone beam computed tomography (kV CBCT) on a linear accelerator (linac) are utilised daily to verify positioning in the multi-week fractionation schedule of a cancer patient undergoing treatment. There is a need for optimisation of radiological protection in kV CBCT imaging protocols to avoid unnecessarily high doses to normal tissues surrounding the target. Traditionally, a 100 mm ionisation chamber and a cylindrical PMMA phantom are used in measuring CT doses, but these are not available in many radiotherapy centres. This study explores an alternative setup using a 0.6 cc Farmer-type ionisation chamber and a 30 cm × 30 cm × 30 cm cubically shaped phantom comprising slabs of water equivalent material typically used in radiotherapy centres. The weighted dose index values determined from the alternative setup are compared with those determined from the conventional cylindrical CT phantoms. The results indicate that the alternative setup using equipment more accessible in radiotherapy facilities is a viable alternative to the use of a CT phantom to quantify kV CBCT radiation dose from linac-based image-guided radiotherapy equipment that can be used for optimisation processes. This should allow most radiotherapy centres all over the world to engage in meaningful dose measurement and optimisation for CBCT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic CT-Based Palliative Radiotherapy Using kV Image Verification: A Prospective Feasibility Study.Current oncology (Toronto, Ont.) · 2026
- Toward Indication-Stratified Local Reference Levels for IGRT Cone-Beam CT: A Clinical Audit of CTDIvol and DLP Across Ten Workflows on Versa HD.Health physics · 2026
- Comparison of extended kV-CBCT image quality between conventional and O-ring linear accelerators.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2026
- Evaluation of quantitative accuracy in 4D DCE‐bCT: Insights from a perfusion phantom studyMedical Physics · 2026
- A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy.Diagnostics (Basel, Switzerland) · 2026
- P10.10.B Optimizing Imaging Dose in Pediatric Neuro-Radiotherapy: Integrating Retrospective Analysis and Predictive Tools for IGRT Dose Reduction within the Image Gently Radiation Safety FrameworkNeuro-Oncology · 2026