P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography

J Matthews, O McGuiness, M Milross, A Piper, K Melehan, F Lowrie, A Myint, C Menadue

Observational

Key points

From the paper's conclusion
  • Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h).

Abstract

Background Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters. Methods A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori. Results Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively). Conclusion Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers