P092. SpO2 agreement between oximeter models and positions
R Cuesta, U Luttrell, E McDermott, E Van Braak, M Spiteri, T Roebuck
Key points
From the paper's conclusion- This study aims to determine SpO2 levels of agreement (LoA) between two oximeter models and explore SpO2 agreement between seated versus lying positions.
Abstract
Commonly, facilities utilise multiple oximeter models that are expected to read similarly. Even if values differ within device specifications it may affect clinical considerations. This study aims to determine SpO2 levels of agreement (LoA) between two oximeter models and explore SpO2 agreement between seated versus lying positions. Data collected at ambulatory sleep study set up. Three values were collected for analysis. T1, SpO2 using Rad-7 (Masimo, Irvine, CA, USA, 2-4sec averaging time, LNCS® cable/sensor) and T2, SpO2 3150 Wrist-Ox2® (Nonin Medical Inc, Plymouth, MN, USA, 3sec averaging time, 8000J sensor). T1 and T2 measurements were taken simultaneously at study setup using same-hand non-thumb digits while seated after 1min stabilisation. T3, average SpO2 during overnight recording for the first 5min artefact-free epoch in a non-upright position scored as wake pre-sleep onset using Noxturnal software (Nox Medical Inc, Reykjavik, Iceland). Primary Bland-Altman analysis, paired t-test and Wilcoxon sign-rank conducted for T1 vs T2 and exploratory analysis for T2 vs T3. Sample n=60 (age 51.6±14.0yrs[24-79], BMI 34.0±8.6kg/m2[19.6-56.0], AHI 25.9±20.9events/hr[1.5-85.4], 32M/28F). T1-T2: bias 1.18% (CI 0.97, 1.40, p<0.001); Wilcoxon signed-rank sensitivity analysis confirmed statistical significance (Z=6.12, p<0.001). LoA -0.45% (CI -0.82, -0.08) to 2.82% (CI 2.44, 3.19). T2-T3 (n=50, 10 excluded for not meeting T3 criteria): bias 0.22% (CI -0.23, 0.67), LoA -2.91 to 3.35%. Small systematic bias observed between oximeter models with agreement narrower than published inter-device studies. No systematic bias between positions but wide individual variation. SpO2, frequently phrased as the fifth vital signal, is impacted by device variation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnographySLEEP advances · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026
- Pulse Oximetry and Arterial Blood Gas Agreement in an Outpatient Black Lung Clinic: Implications for Clinical Assessment and Decision Making.Journal of primary care & community health · 2026
- Sleep position, respiratory event rates, and nocturnal oxygenation in a clinical monitoring cohort: a positional phenotyping analysis.Frontiers in neurology · 2026
- Pulse Oximeter Performance and Skin Pigment: Comparison of 34 Oximeters Using Current and Emerging Regulatory FrameworksAnesthesia & Analgesia · 2026
- Effect of lingual gauze swab placement and light exclusion on pulse oximeter agreement with arterial haemoglobin-oxygen saturation derived from blood-gas analysis in healthy anaesthetised horses.Veterinary anaesthesia and analgesia · 2026