Performance Inconsistency of Pulse Oximeters Over Five Years of Clinical Use: A Single-Center Study of Masimo Devices.
Ashraf Fawzy, Michelle M Qin, Xinyi Sun, Li Yan, Chad H Hochberg, Peter H Dziedzic, Jose I Suarez, Yanxun Xu,
Division of Pulmonary and Critical Care, Department of Medicine, Johns Hopkins University School of Medicine, Baltimore, MD, USA. · Department of Biostatistics, Johns Hopkins University Bloomberg School of Public Health, Baltimore, MD, USA.
Key points
From the paper's conclusion- The accuracy of the tested medical-grade pulse oximeter changed year-to-year at a large hospital compared to concurrent arterial blood gas saturation, potentially complicating safe clinical decision-making and application of evidence-based medicine. Errors of false hypoxemia became more common, threatening different clinical consequences and necessitating different adaptations than errors leading to occult hypoxemia.
Abstract
RATIONALE: Clinical recognition and treatment of hypoxemia rely on pulse oximetry. Despite general awareness of pulse oximeter shortcomings, clinicians may assume consistent performance across time, models, and batches, an assumption essential for applying clinical trial evidence and guidelines at the bedside. OBJECTIVES: Assess whether the difference between the estimated arterial oxygen saturation from a medical-grade Masimo pulse oximeter versus gold-standard concurrent arterial blood gas changed over 5 years of clinical use in intensive care unit (ICU) patients. METHODS: Patients with ≥1 concurrent arterial oxygen saturation (SaO2) and pulse oximetry (SpO2) in an ICU with high-frequency vital sign data (SpO2 recorded at 2-second intervals) from 2020-2024 were included. Occult hypoxemia (SaO2<88% with SpO2≥92%), and false hypoxemia (SpO2<88% with SaO2≥92%) were examined by calendar year, as were mean oximeter error (SpO2-SaO2), Accuracy Root Mean Square (Arms, the U.S. Food and Drug Administration (FDA)'s preferred measure), and errors >4% and <-4%. Results were stratified by race/ethnicity as recorded in the electronic medical record. An adjusted linear mixed effects model evaluated calendar-year changes. MEASUREMENTS AND MAIN RESULTS: 223,145 ABGs in 15,129 patients were analyzed. Occult hypoxemia and overestimation errors >4% decreased while false hypoxemia and underestimation errors <-4% increased several-fold between 2020-2024. Mean oximeter error shifted from overestimation to underestimation by mid-2022. Arms fluctuated and was consistently above 3-3.5%, the FDA's threshold for controlled laboratory testing among healthy volunteers, peaking at end of study. Fluctuation was greatest for SaO2<92% and among non-White patients. By 2024, oximeters underestimated oxygen saturation on average in adjusted models for both Black and White patients with SaO2≥92% and showed relative underestimation among Black versus White patients across SaO2 categories. CONCLUSIONS: The accuracy of the tested medical-grade pulse oximeter changed year-to-year at a large hospital compared to concurrent arterial blood gas saturation, potentially complicating safe clinical decision-making and application of evidence-based medicine. Errors of false hypoxemia became more common, threatening different clinical consequences and necessitating different adaptations than errors leading to occult hypoxemia.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The impact of skin tone on performance of pulse oximeters used by NHS England COVID Oximetry @home scheme: measurement and diagnostic accuracy study.BMJ (Clinical research ed.) · 2026
- Occult hypoxemia in the ICU: Clinical risk factors, mortality and implications for nursing practice.Intensive & critical care nursing · 2026
- Pulse Oximeter Performance and Skin Pigment: Comparison of 34 Oximeters Using Current and Emerging Regulatory FrameworksAnesthesia & Analgesia · 2026
- P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnographySLEEP advances · 2026
- Pulse Oximetry and Arterial Blood Gas Agreement in an Outpatient Black Lung Clinic: Implications for Clinical Assessment and Decision Making.Journal of primary care & community health · 2026
- P092. SpO2 agreement between oximeter models and positionsSLEEP advances · 2026