Pulse Oximeter Performance and Skin Pigment: Comparison of 34 Oximeters Using Current and Emerging Regulatory Frameworks
Caroline Hughes, Danni Chen, Tyler J. Law, Philip E. Bickler, John Feiner, Leonid Shmuylovich, Ella Behnke, Lily Ortiz,
University of California, San Francisco · Washington University in St. Louis
Key points
From the paper's conclusion- We tested common oximeters with current and anticipated regulatory frameworks.
Abstract
BACKGROUND: The International Organization for Standardization (ISO) and US Food and Drug Administration (FDA) are updating regulations for pulse oximeters to reduce performance disparities linked to skin pigment. We tested common oximeters with current and anticipated regulatory frameworks. We hypothesized that not all oximeters show more positive bias in darkly vs lightly pigmented participants and that few oximeters would "pass" the anticipated FDA regulations. METHODS: We used a controlled desaturation protocol to test 34 oximeters across arterial oxygen saturations (SaO2) 70% to 100% in healthy adults. Based on what FDA and ISO had shared at the time of study design, we studied cohort sizes of ≥24 with ≥25% of participants being darkly pigmented. We used the subjective Monk Skin Tone (MST) scale and the objective individual typology angle (ITA) derived from a spectrophotometer to characterize skin pigment. The accuracy root mean square error (ARMS), bias (mean of SpO2 - SaO2 error), and skin pigment differential bias were calculated. Monte Carlo simulation explored potential impacts of participant selection on device passing. RESULTS: For cohorts of 24 participants, 28/34 oximeters (82%) passed 2017 ISO Standard (ARMS < 4%), 22/34 (65%) passed 2013 FDA guidance (ARMS ≤ 3%), 21/34 (62%) oximeters passed both ARMS and differential bias criteria for anticipated ISO standards, and 1/34 (3%) passed anticipated FDA criteria. More devices passed when testing cohorts were increased beyond 24 participants. Eleven oximeters had more positive bias in participants with dark vs light (dorsal finger) pigmentation across 70% to 100% SaO2 (median difference 1.42, IQR: 1.35-1.90). Eighteen devices could pass or fail depending on cohorts selected for analysis. CONCLUSIONS: Pulse oximeters show variable performance across manufacturers and models. Notably, only some devices show more positive bias in people with darker skin. Anticipated updates to ISO and FDA frameworks yield strikingly different assessments and require refinement of cohort sizes and differential bias criteria. Whether new guidelines will translate into improved real-world performance or reduced health disparities is yet to be determined.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The impact of skin tone on performance of pulse oximeters used by NHS England COVID Oximetry @home scheme: measurement and diagnostic accuracy study.BMJ (Clinical research ed.) · 2026
- Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation.Frontiers in digital health · 2026
- P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnographySLEEP advances · 2026
- Performance Inconsistency of Pulse Oximeters Over Five Years of Clinical Use: A Single-Center Study of Masimo Devices.Annals of the American Thoracic Society · 2026
- Effects of Skin Pigmentation on Pulse Oximetry: A Pilot Study.American journal of critical care : an official publication, American Association of Critical-Care Nurses · 2026
- Pulse Oximetry and Arterial Blood Gas Agreement in an Outpatient Black Lung Clinic: Implications for Clinical Assessment and Decision Making.Journal of primary care & community health · 2026