A-412 To hydrate or not to hydrate? A comparison of the Siemens epoc, Abbott i-STAT, and Beckman Coulter DxC700AU IDMS traceable creatinine assays’ inter-platform variability
Michelle Zimmerman, Susie Chambers, Nicole Therien, Melissa Jordan
Key points
From the paper's conclusion- The Siemens epoc exhibited a consistent negative bias relative to the DxC700AU, while the Abbott i-STAT showed a positive bias.
Abstract
Background Indiana University Health (IUH) transitioned from the Abbott i-STAT system to the Siemens epoc platform to modernize point of care (POC) testing. After implementation, radiology departments at multiple IUH hospitals reported lower creatinine (Cr) and higher eGFRcr values on epoc compared to i-STAT. These differences affected hydration decisions for contrast media imaging, prompting a system wide investigation. We sought to evaluate Cr result differences across platforms and determine potential clinical impacts, especially around key medical decision points influencing imaging protocols. Methods This study was conducted across fourteen IU Health hospitals between July 29 and August 30, 2024. Each hospital was assigned a specified number of samples to obtain based on POC testing volumes at that location. The Beckman Coulter DxC700AU laboratory analyzer served as the reference method. Both POC devices use enzymatic creatinine measurement methodologies while the laboratory analyzer uses a kinetic modification of the Jaffe procedure. The 2021 CKD-EPI equation with no race modifier was used for eGFRcr calculation. Specimens were collected from patients needing intravenous access for various procedures, including dialysis, who were also having labs drawn at the same time. Specimens were tested on both POC devices and then centrifuged for testing on the DxC700AU. Results were assessed using EP Evaluator’s multi-instrument comparison module. For each analyte, the target value was defined as the median result generated across all three platforms. Agreement between individual platform results and the target value was evaluated according to predefined Total Allowable Error limits of ±10% or ±0.2?mg/dL, whichever was greater. To assess the DxC700AU’s performance on commutable material, the College of American Pathologists’ LN24 Creatinine Accuracy Calibration Verification/Linearity program (2024 LN24-B) samples were used. Results Specimens from 418 adult patients were collected. Patients ranged from 19 to 93 years with a mean age of 65±14 years. Males accounted for 50.7%. The range of Cr values for the epoc was 0.34-12.91 mg/dL; for the i-STAT, 0.20-13.70 mg/dL, and for the DxC700AU, 0.25-12.96 mg/dL. Approximately 80% of the Cr values = 1.5 mg/dL. The LN24 results analyzed on the DxC700AU gave Cr results similar to those published in the corresponding Participant Summary. Distinct directional bias for Cr was seen between the POC platforms when compared against our reference method. The Siemens epoc exhibited a consistent negative bias relative to the DxC700AU, while the Abbott i-STAT showed a positive bias. When evaluating the potential clinical implications, 23 of the 418 samples (5.5%) would have resulted in different patient triage decisions for our health system’s eGFRcr based contrast media hydration protocol. Protocol agreement between DxC700AU and epoc eGFRcr was 11/23 (48%), for DxC700AU and i-STAT, 12/23 (52%), and for epoc and i-STAT, 0/23 (0%). Conclusion The effort to harmonize Cr assays began over 20 years ago. The three instruments in this study use IDMS traceable Cr methodologies but their results indicate harmonization could still be improved. Although a small percentage of Cr values would have resulted in a patient being triaged differently, this study underscores the potential clinical impact of result divergence among platforms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-411 Evaluation of Point-of-Care Creatinine Analyzers: eGFR Categorization Accuracy for Safe Contrast AdministrationClinical Chemistry · 2026
- A-002 Evaluation of the Abbott i-STAT devices for both high-sensitivity cardiac troponin I and creatinineClinical Chemistry · 2026
- Evaluation of Point-of-Care Testing for Creatinine, Estimated Glomerular Filtration Rate, and Potassium in a Diverse United Kingdom Population.The journal of applied laboratory medicine · 2026
- B-288 Comparison of Three CKD-EPI eGFR Equations for the Assessment of Kidney DiseaseClinical Chemistry · 2026
- A-399 Analytical performance of enzymatic amperometric glucose and creatinine sensors for the Fluxergy multimodal point-of-care platformClinical Chemistry · 2026
- B-274 Performance Evaluation of Direct ISE Method and Methodological Comparison Among Multiple Detection PlatformsClinical Chemistry · 2026