Extending Quality Control Procedures for Multiplexed Clinical Assays.
Daisy Unsihuay, Stephen R Master
Department of Pathology and Laboratory Medicine, Henry Ford Hospital, Detroit, MI, United States. · Department of Pathology and Laboratory Medicine, The Children's Hospital of Philadelphia, Philadelphia, PA, United States.
Key points
From the paper's conclusion- These results demonstrate multiple QC strategies that control highly multiplexed assays (1000-plex) at a level that is comparable to traditional QC schemes for single-analyte assays.
Abstract
BACKGROUND: Multiplexed assays, which simultaneously measure many analytes from a single sample, have become increasingly significant for laboratory diagnosis. Many multiplexed assays with application to clinical diagnosis consist of analyte measurements that can fail individually. In these cases, traditional statistical quantitative quality control (QC) measures cannot be used without creating an unacceptably high false rejection rate. METHODS: We developed stochastic simulation software (qcsim) to calculate and visualize the detection power of complex QC rule combinations, including traditional Westgard rules as well as statistical tests of multiple QC repeats, with arbitrary degrees of multiplexing and levels of control. We used this approach to evaluate novel QC models that maintain stringent control of the bias and imprecision of each analyte in a highly multiplexed assay ("panel"). For classifier-based assays that use an algorithm to generate a small number of diagnostic outcomes from a large number of analytes (a "pattern"), we use perturbation analysis to assess the effect of different classifiers with a single analytical platform. RESULTS: Multiple QC approaches are able to overcome the challenge of highly multiplexed assays, and we demonstrate successful strategies that control the false rejection rate using either high analytical performance (low imprecision) or multiple QC replicates. We also demonstrate that, for pattern-based assays, algorithmic details of the specific classifier determine both critical analytes and the required stringency of the QC design. CONCLUSIONS: These results demonstrate multiple QC strategies that control highly multiplexed assays (1000-plex) at a level that is comparable to traditional QC schemes for single-analyte assays.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Three consecutive failures, one uncertain verdict: A simulation-based reassessment of accreditation-restriction rules in clinical chemistryRevista Romana de Medicina de Laborator · 2026
- Application of Patient-Based p Control Chart for Quality Control in a Clinical Laboratory.Annals of laboratory medicine · 2026
- A-275 Estimating the Number of Erroneous Results due to an Out-of-Control ConditionClinical Chemistry · 2026
- Clustering of patient-derived laboratory data: A privacy-preserving distributed approach to enhance clinical laboratory quality control and regional external quality assessment.PloS one · 2026
- Accurate quantification in proteomics with QuantUMSNature Biotechnology · 2026
- Content uniformity testing with PAT tools- how does the probability of batch acceptance change?International journal of pharmaceutics · 2026