Three consecutive failures, one uncertain verdict: A simulation-based reassessment of accreditation-restriction rules in clinical chemistry
Radu Ilinca, Dan Adrian Luțescu, Ionela Ganea, Daniela Cristina Dimitriu
Key points
From the paper's conclusion- Conclusions Three consecutive unsatisfactory PT results are not an invariant indicator of deficient analytical performance.
Abstract
Background Proficiency testing (PT) is central to laboratory quality assurance; however, repeated unsatisfactory results may reflect both analytical performance and PT-scheme architecture. We evaluated whether three consecutive unsatisfactory results reliably identify deficient analytical performance in serum creatinine testing. Methods Monte Carlo simulation modelled three-round trajectories in an ISO 13528-based PT scheme. A total of 206 prespecified scenarios generated 1,446 analytical conditions varying laboratory performance, peer-group definition and size, heterogeneity, consensus construction, between-round persistence and σPT derivation. More than 875 million three-round trajectories were simulated. False-trigger and detection probabilities were evaluated for competent profiles and deficiencies caused by systematic bias, increased analytical imprecision or both. Results Under the reference condition, false-trigger probabilities were 0-0.03% for competent profiles, reaching 2.81% under the most adverse configuration examined. Detection probabilities were 20.47% for systematic bias, 0.38% for increased analytical imprecision and 16.24% for combined deficiency. With analytical performance held constant, changes in peer-group definition, focal-laboratory contribution to consensus or σPT derivation altered trigger probabilities by approximately 1.3- to 1.9-fold. Full between-round persistence increased false-trigger probability by approximately 24% relative to independence. Distributional sensitivity analysis produced a maximum absolute difference of 0.69 percentage points. Conclusions Three consecutive unsatisfactory PT results are not an invariant indicator of deficient analytical performance. Their probability depends on both the mechanism of deficiency and PT-scheme architecture. Such outcomes should therefore trigger investigation rather than independently establish analytical deficiency or its cause.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Qualitative results in proficiency testing: a systematic review of evaluation practices and error integration.Clinical chemistry and laboratory medicine · 2026
- A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.Diagnostics (Basel, Switzerland) · 2026
- A-275 Estimating the Number of Erroneous Results due to an Out-of-Control ConditionClinical Chemistry · 2026
- B-281 Comparison of Analytical Error Frequencies Between Closed-System and Open-System Clinical Chemistry AnalyzersClinical Chemistry · 2026
- Quantifying and adjusting for verification bias in diagnostic test accuracy studies with conditional reference standardsBMC Medical Research Methodology · 2026
- Assigned-Value Uncertainty in Limited-Participant Proficiency Testing: Monte Carlo Evaluation of ISO 13528 Estimators in an Industrial Calibration ScenarioMetrology · 2026