Clustering of patient-derived laboratory data: A privacy-preserving distributed approach to enhance clinical laboratory quality control and regional external quality assessment.
Chao Song, Qian Chen, Chunsheng Qu, Liqing Xu, Zhiming Shan, Jun Zhou, Weixing Li
Laboratory Medicine Center, Zhejiang Center for Clinical Laboratories, Zhejiang Provincial People's Hospital, Affiliated People's Hospital, Hangzhou Medical College, Hangzhou, Zhejiang, China. · Department of Cilinical laboratory, Lishui Hospital of Wenzhou Medical University, The First Affiliated Hospital of Lishui University, Lishui People's Hospital, Lishui, Zhejiang, China.
Key points
From the paper's conclusion- Density-based clustering underpins the PBRTQC-PBQA model's effectiveness in deriving hospital-specific core reference values, filtering unconfirmed QC alerts, validating calibration, and enhancing regional test consistency. The model's ability to detect systematic biases not captured by IQC underscores its value as a complementary quality assurance layer. Its lightweight design supports regional deployment.
Abstract
BACKGROUND: Traditional quality control (QC) has inherent limitations, and there are still various challenges existing in patient-based real-time QC (PBRTQC) and patient-based quality assurance (PBQA). We developed an integrated PBRTQC-PBQA model using density-based clustering and evaluated its effectiveness as a regional clinical laboratory quality assurance tool. METHODS: A 20-month multicentre study across seven hospitals analysed five haematological analytes(white blood cell count [WBC], red blood cell count [RBC], haemoglobin [HGB], platelet count [PLT], and mean corpuscular volume [MCV]). Custom Clustering-PBRTQC-PBQA(CPP) models were developed using log/Box-Cox transformation, 3D spatial construction with companion analytes determined by distribution morphology, Mahalanobis distance filtering (chi-square = 0.3, ~ 47% exclusion rate), and iterative OPTICS clustering to identify hospital-specific core reference populations. A privacy-preserving distributed architecture transmitted only aggregated statistical summaries. Median/EWMA harts (PBRTQC) and Z-score/relative bias charts (PBQA) were created alongside a closed-loop quality improvement system and calibration event verification. RESULTS: The CPP model reduced EWMA alert episodes by 25.0% (92-69) while increasing 1-3S alert episodes by 31.2% (64-84) compared with conventional moving-average PBRTQC, demonstrating selective elimination of systematic bias signals while preserving sensitivity to random errors (net change: 180-173 episodes, -3.9%). Overall calibration consistency was 82.3% (95% CI: 74.1%-88.7%). Among 52 confirmed true-positive alerts, 37 (71.2%) were detected without concurrent internal QC out-of-control signals, demonstrating the model's capacity to capture low-magnitude biases that conventional IQC misses. CONCLUSION: Density-based clustering underpins the PBRTQC-PBQA model's effectiveness in deriving hospital-specific core reference values, filtering unconfirmed QC alerts, validating calibration, and enhancing regional test consistency. The model's ability to detect systematic biases not captured by IQC underscores its value as a complementary quality assurance layer. Its lightweight design supports regional deployment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Application of Patient-Based p Control Chart for Quality Control in a Clinical Laboratory.Annals of laboratory medicine · 2026
- Leveraging Patient Data for Real-Time Inter-Instrument Comparison: Performance of the EWMA-Based PBRTQC Model on Triglyceride Measurements.Journal of clinical laboratory analysis · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026
- Illustrative examples and data from patient-based real-time quality controlThe Clinical Biochemist Reviews · 2026
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.Diagnostics (Basel, Switzerland) · 2026