A-275 Estimating the Number of Erroneous Results due to an Out-of-Control Condition
David Bauer, Alex Jafek, John Yundt-Pacheco
Key points
From the paper's conclusion- Conclusion This abstract presents a way to estimate the number of compromised patient specimens due to an Out-of-Control condition and how to estimate the probability of an individual patient specimen being compromised by a given error condition.
Abstract
Background Dealing with the aftermath of a confirmed Out-of-Control Condition is a critical part of clinical diagnostic laboratory testing. After troubleshooting and correction steps are completed, decisions need to be made about which patient specimens need to be re-evaluated, and which need to be corrected. In the absence of additional information, the conservative approach is to retest all patient specimens that were tested after the last accepted quality control evaluation. This much retesting may present a significant logistical challenge that requires unplanned time and cost. An alternative approach is to estimate the number of patient specimens that were compromised by the Out-of-Control condition and direct the recovery accordingly. The method presented below makes this estimation by first characterizing Out-of-Control condition bias, imprecision, and TEa curves based on analyte concentration. The probability of each specimen being outside its quality specification is computed and summed to estimate the number of patient specimens compromised. Methods To collect the data needed to characterize the Out-of-Control condition, each level of QC is run in triplicate, so there is a total of 4 datapoints (with the original QC evaluation that was Out-of-Control). Bias is computed by comparing the Out-of-Control means with the In-Control QC means. The bias, TEa curves are concentration based. For concentrations at or below the lowest QC concentration, the bias, imprecision and TEa of the lowest QC mean is used. For concentrations at or above the highest QC concentration, imprecision and TEa of the highest QC mean is used. For concentration between QC means, interpolation is used to assign concentration specific bias, and TEa. The probability of a specimen result x being outside its quality specification (Unreliable) is computed as: pU(x) = 1 – normcdf (TEa(x), bias(x), sd(x)) + normcdf(-TEa(x), sd(x)) where TEa(x), sd(x) are the concentration- based bias, and TEa curves. The normcdf(x) is the Normal cumulative distribution function. The sum of the pU(x) probabilities is the estimated number of compromised patient specimens. Results Using 2015 patient results from a GGT test with a TEa=5 for 40 U/L, QC means of 37.14 and 153.0, QC SDs of 0.81 and 3.45, the estimated number of erroneous results were computed with error sizes ranging from -10SDs to +10SDs. The sigma metrics for the GGT test range from 6.17 (low concentration) to 5.32 (high concentration). The %Unreliable Results is plotted against the size of the error condition from -10SDs to +10SDs (see figure). It takes 4.20SDs to compromise 5% of the patient results. 6.03SDs to compromise 50%, and 9.38SDs to compromise 95% of the patient results. Conclusion This abstract presents a way to estimate the number of compromised patient specimens due to an Out-of-Control condition and how to estimate the probability of an individual patient specimen being compromised by a given error condition. The relationship between the size of an Out-of-Control condition and the % of the patient results compromised is not linear until the Out-of-Control condition is so large that it compromises every patient specimen being tested.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Application of Patient-Based p Control Chart for Quality Control in a Clinical Laboratory.Annals of laboratory medicine · 2026
- Three consecutive failures, one uncertain verdict: A simulation-based reassessment of accreditation-restriction rules in clinical chemistryRevista Romana de Medicina de Laborator · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026
- A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.Diagnostics (Basel, Switzerland) · 2026
- A-282 Analysis of Recheck and Revision Rates Across the Life Cycle of High Volume Chemistry Analyzers: Informing Lab Stewardship and Asset ManagementClinical Chemistry · 2026
- Improving Internal Quality Control Through Metrological Reinterpretation of Westgard RulesMetrology · 2026