Application of Patient-Based p Control Chart for Quality Control in a Clinical Laboratory.
Tingting Long, Jiali Liu, Wanzhou Xu, Pingan Zhang, Yuzhu Huang, Zhiguo Wang
Department of Clinical Laboratory, Renmin Hospital of Wuhan University; Institute of Clinical Molecular Diagnostics, Wuhan University; Hubei International Science and Technology Cooperation Base for Digital and Intelligent Molecular Diagnostics, Wuhan, Hubei, China. · National Center for Clinical Laboratories, Beijing Hospital, National Center for Gerontology; National Clinical Research Center for Gerontology; The Key Laboratory of Geriatrics of NHC; Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, China.
Key points
From the paper's conclusion- The patient-based p control chart complemented laboratory QC analysis, with performance influenced by positive rate, threshold location, and subgroup size.
Abstract
BACKGROUND: Patient-based QC is a critical complementary tool for laboratory quality management. We evaluated the application and statistical performance of a patient-based p (a proportion-based statistical control chart) across quantitative and qualitative laboratory assays. METHODS: We collected 413,586 valid test records involving 10 laboratory assays and divided them into training and verification datasets. P control charts were constructed using daily positive proportions and sample-size-specific limits. Monte Carlo simulations evaluated in-control performance, shift-detection performance, and sensitivity to group size using theoretical binomial shifts, empirical bootstrap simulations with analytical bias, and fixed subgroup sizes. RESULTS: Empirical false-positive rates under in-control conditions ranged from 0.25% to 0.50%. Detection power increased with the magnitude of event-proportion or analytical shifts, whereas directional asymmetry occurred for thresholds located in distribution tails. Sample-size simulations showed greater stabilization of false-positive rates at daily subgroup sizes of ≥ 200. With clinical monitoring, the p control chart identified sustained increases in HbA1c positivity exceeding the diagnostic cut-off and detected persistent changes in total protein patient-result proportions that were not accompanied by conventional internal QC violations. Clinical decision limits-based charts for Na and K generated fewer signals than did reference intervals-based charts. In HBsAg monitoring, the p control chart remained within control limits following reagent lot replacement, whereas the corresponding Levey-Jennings charts showed an approximately 4-SD shift. CONCLUSIONS: The patient-based p control chart complemented laboratory QC analysis, with performance influenced by positive rate, threshold location, and subgroup size.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clustering of patient-derived laboratory data: A privacy-preserving distributed approach to enhance clinical laboratory quality control and regional external quality assessment.PloS one · 2026
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026
- A-275 Estimating the Number of Erroneous Results due to an Out-of-Control ConditionClinical Chemistry · 2026
- Patient-Based Real-Time Quality Control for Semiquantitative Tests Based on Hierarchical Differences.Annals of laboratory medicine · 2026
- Illustrative examples and data from patient-based real-time quality controlThe Clinical Biochemist Reviews · 2026