B-387 Analytical Precision and on-board stability Evaluation of the EMIT Methotrexate and Amikacin and EMIT 2000 Vancomycin and Digoxin Assays on the Atellica DT 250 Analyzer
Jessica Daughtry, Kenisha Griffin, Jerry Glover, Nicholas Johnson, Benjamin Israel
Key points
From the paper's conclusion- Conclusion The EMIT Methotrexate, Vancomycin, Amikacin, and Digoxin assays on the new Atellica DT 250 Analyzer demonstrated low CVs, strong correlation, and precise, robust performance across clinically relevant concentrations.
Abstract
Background Therapeutic drug monitoring (TDM) is the measurement of the serum level of a drug to ensure its concentration is within the therapeutic range. The EMIT Methotrexate (MTX), Amikacin (AMIK), Vancomycin (VANC), and Digoxin (DIG) assays are homogeneous enzyme immunoassays that can be used to aid in the diagnosis and treatment of overdose of these drugs as well as monitoring levels for therapeutic administration. Protocols for these assays have been developed on the new Atellica DT 250 Analyzer. The data presented here were generated on the Atellica DT 250 Analyzer. Methods Precision was evaluated at three levels for each assay. Each assay’s range was evaluated by spiking drug at various levels into human serum. Specimens were analyzed and the results were compared to those of the Viva-E/Viva-ProE Drug Testing Systems. On-instrument stability was assessed by testing the assay controls over time. Results Precision evaluation demonstrated repeatability and within-in lab CVs ranging from 1.8 to 4.0% for MTX, 2.6 to 5.6% for AMIK, 1.2 to 5.9% for VANC, and 4.0 to 11.8% for DIG. The assays quantified drug-spiked samples for sample concentrations between 0.3 and 2.0 µmol/L for MTX, 2.5 and 50.0 µg/mL for AMIK, 2.0 and 50.0 µg/mL for VANC, and 0.30 and 5.00 ng/mL for DIG. Specimens run on the Atellica DT 250 Analyzer and Viva-E/Viva-ProE Systems had a slope of 0.95-1.08 and r = 0.95 for all assays. On-instrument stability for all assays on the Atellica DT 250 Analyzer was found to be up to 8 weeks. Conclusion The EMIT Methotrexate, Vancomycin, Amikacin, and Digoxin assays on the new Atellica DT 250 Analyzer demonstrated low CVs, strong correlation, and precise, robust performance across clinically relevant concentrations. In addition, the analyzer supported up to eight weeks of onboard reagent stability for all assays.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 132 Improving Phenytoin Therapeutic Drug Monitoring Accuracy Using Six Sigma MetricsAmerican Journal of Clinical Pathology · 2026
- B-392 Clinical Impact of DAMPA Interference and Low Concentration Performance of the Roche Methotrexate ImmunoassayClinical Chemistry · 2026
- A pH compensation and peak identification algorithm for voltammetric measurement of therapeutic drugs with sweat sensorsTalanta · 2026
- Inter-laboratory variability in vancomycin measurement in an external quality assessment scheme 2016–2025Clinical Biochemistry · 2026
- Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.Journal of pharmaceutical health care and sciences · 2026
- 6 Validation of a Dilute Thrombin Time Assay for Pediatric Patients on Bivalirudin TherapyAmerican Journal of Clinical Pathology · 2026