B-392 Clinical Impact of DAMPA Interference and Low Concentration Performance of the Roche Methotrexate Immunoassay
Mehdi Hooshmandi, Jennifer Shea, Raymond Ngo, Julie Podesky, Bonny Lem Ragosnig, Swati Swati, Chien-Yi Lu, Lei Fu,
Key points
From the paper's conclusion- Conclusion The Roche MTX immunoassay exhibits marked and prolonged positive interference from DAMPA after glucarpidase, resulting in substantial MTX overestimation compared with LC-MS/MS.
Abstract
Background Methotrexate (MTX) therapeutic drug monitoring is essential after high-dose therapy, particularly at low concentrations where clinical decision thresholds guide rescue interventions. Glucarpidase rapidly hydrolyzes MTX to DAMPA, which strongly cross-reacts with MTX immunoassays and produces falsely elevated results. Roche Diagnostics recently released a new MTX immunoassay; its analytical performance and susceptibility to DAMPA interference require evaluation to determine clinical suitability. We assessed assay accuracy and precision and compared results with other commercial immunoassays and LC-MS/MS using post-glucarpidase patient samples. Methods Serial serum samples were collected from two glucarpidase-treated patients over 5–11 days. At each time point, MTX concentrations were measured in parallel using a LC-MS/MS and three automated immunoassays: Abbott Architect Methotrexate assay, Roche ONLINE TDM Methotrexate assay on cobas® c503, and ARK™ Methotrexate assay on cobas® c503. Percent differences between each immunoassay and LC-MS/MS were calculated and trended to assess DAMPA-related interference and to compare assay susceptibility. To assess analytical precision, a between-day imprecision study over 13–21 days was conducted at three laboratory sites using three levels of Roche quality control (QC) material (target concentrations: 0.08, 0.49, and 0.80 µmol/L for levels 1, 2, and 3, respectively). The coefficient of variation (CV%) was calculated for each QC level. A serum pool with MTX at approximately 0.1 µmol/L concentration was also prepared and sent to the sites for precision evaluation as well. Results In both patients, all three immunoassays markedly overestimated MTX concentrations compared with LC-MS/MS following glucarpidase administration. In Case 1, peak positive bias occurred at 24 hours, reaching 36,700% (Roche), 32,900% (ARK), and 13,500% (Architect). Although bias declined over time, substantial overestimation persisted through 120 hours (93–122% for Roche/ARK vs 12% for Architect). In Case 2, similar patterns were observed, with extreme positive bias at 48 hours (4,322% ARK; 3,511% Roche; 1,233% Architect), decreasing but remaining clinically significant at 120 hours (130–131% for ARK/Roche vs 20% for Architect). For both cases, the Abbott Architect assay demonstrated the lowest susceptibility to DAMPA interference, whereas Roche and ARK exhibited greater magnitude and prolonged positive bias. Between-day precision demonstrated excellent analytical performance at QC levels 2 and 3 in Roche MTX assay, with CVs <2% across three study sites. In contrast, imprecision at the low-level control was significantly higher, with CVs ranging from 7.7% to 12.56%. This pattern was consistently observed across all participating sites. To maintain the CV for QC level 1 within predefined acceptability criteria, frequent recalibration was required to restore control results to the assigned target range. Serum pool also exhibited a downward trend with daily testing over 4-5 days. This imprecision at low concentration was not observed in the other assays. Conclusion The Roche MTX immunoassay exhibits marked and prolonged positive interference from DAMPA after glucarpidase, resulting in substantial MTX overestimation compared with LC-MS/MS. Accurate post-glucarpidase monitoring requires LC-MS/MS. In addition, calibration drift at low MTX concentrations highlights the need for close QC monitoring and timely recalibration to ensure reliable interpretation of clinical decision cut-offs near 0.1 µmol/L.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Differential Electrochemical Sensor for Resolving Metabolite Cross-Reactivity Toward Selective Chemotherapy Monitoring.ACS sensors · 2026
- B-387 Analytical Precision and on-board stability Evaluation of the EMIT Methotrexate and Amikacin and EMIT 2000 Vancomycin and Digoxin Assays on the Atellica DT 250 AnalyzerClinical Chemistry · 2026
- Drivers of Methotrexate Polyglutamate Concentration in Erythrocytes: Insights from Immune-Mediated Inflammatory Diseases and Pediatric Acute Lymphoblastic LeukemiaPharmaceuticals · 2026
- Resolving Therapeutic Monoclonal Antibody Interference in an Immunoglobulin M Kappa Multiple Myeloma Patient Using a Mass Spectrometry Assay.Clinical laboratory · 2026
- Fixed-dose glucarpidase use for high-dose methotrexate toxicity: A case series in Acute Lymphoblastic Leukemia.Journal of oncology pharmacy practice : official publication of the International Society of Oncology Pharmacy Practitioners · 2026
- Methotrexate toxicity due to a medication dispensing error compounded by an interaction with proton pump inhibitorsToxicology Reports · 2026