6 Validation of a Dilute Thrombin Time Assay for Pediatric Patients on Bivalirudin Therapy

Diana Arroyo Lugo, Morayma Reyes Gil

Key points

From the paper's conclusion
  • Conclusion The DTT is more reliable, consistent and accurate than the aPTT at monitoring pediatric patients on bivalirudin therapy with significantly less interferences and physiological variables.

Abstract

Background and Aim Measurement of direct thrombin inhibitors (DTIs) can be challenging for the coagulation laboratory and the clinical team monitoring anticoagulation. Several studies have demonstrated that the dilute thrombin time (DTT) assay is a more robust assay for measuring DTI levels than the traditional activated partial thromboplastin time (aPTT). The DTT test is a simple assay where patient plasma is diluted with a normal pool 1:4 before performing a thrombin time test. Pediatric patients on extracorporeal membrane oxygenation (ECMO) or ventricular assist devices (VAD) are frequently anticoagulated using bivalirudin, a DTI. The aim was to validate a DTT assay to provide measurements of bivalirudin in these critical patients. Methods 295 serial samples collected for bivalirudin monitoring using the aPTT from 8 pediatric patients were also tested for DTT using the clot-based detection instrumentation. To establish the reference range for DTT, 25 normal donor samples were tested, and the normal range was established at < 18.6s. Accuracy, precision, and system correlation data was gathered from 3 identical instruments. In addition, interferences from low and high fibrinogen levels, high D-dimer, and gross hemolysis were tested with 10 samples each. Results For inter- and intra-precision, all quality control data was valid, and the CV was within our target of 10%. For the system correlation, 20 samples were run, and all had CVs within our target of 20%. No interferences were found from low or high fibrinogen or from high D-dimer results. Only samples that reached 4+ hemolysis grading sometimes saw a 15% decrease in the DTT. Some patients on bivalirudin never became therapeutic on either aPTT (60-80s) or DTT (60-90s). Of those that did have therapeutic aPTTs (n = 61), 59.02% (n = 36) also had therapeutic DTTs. This illustrates the difficulty of monitoring bivalirudin using only the aPTT. Moreover, when using a calibrated DTT assay, based on the bivalirudin dose-response curve with bivalirudin calibrators, only 29.51% of the therapeutic PTT samples had therapeutic bivalirudin concentrations (1.23-2.46 mcg/mL) for patients on ECMO, based on a retrospective quality improvement project. Finally, of the 8 original pediatric patients, 5 seem therapeutic with both the aPTT and DTT assays. 3 were found to have conflicting DTT and aPTT results; 2 were subtherapeutic by DTT with “therapeutic” aPTT results, and 1 was in the high therapeutic range by DTT but with” subtherapeutic” aPTT results. Conclusion The DTT is more reliable, consistent and accurate than the aPTT at monitoring pediatric patients on bivalirudin therapy with significantly less interferences and physiological variables. The DTT test was successfully validated for the laboratory and currently available to monitor patients on DTIs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

6 Validation of a Dilute Thrombin Time Assay for Pediatric Patients on Bivalirudin Therapy — Science Skim