Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.
Takashi Ohtsuki, Yume Enomoto, Akane Yasuda, Risa Imai, Nonoka Seto, Makoto Hashimoto, Miyuki Matsushita, Satoru Morikawa,
College of Bioresource Sciences, Nihon University, 1866 Kameino, Fujisawa, Kanagawa, Japan. ohtsuki.takashi@nihon-u.ac.jp. · College of Bioresource Sciences, Nihon University, 1866 Kameino, Fujisawa, Kanagawa, Japan.
Key points
From the paper's conclusion- The validated RMS-based HPLC method provides accurate and precise quantification of structurally diverse drugs using stable, the International System of Units (SI)-traceable, non-analyte reference materials. This approach eliminates the need for analyte-specific authentic standards for routine analyses, thus substantially reducing procurement costs and addressing the instability of some analytes. The rapid and accurate analysis enables timely reporting, potentially leading to improved clinical outcomes through expedited in-house TDM implementation in hospital laboratories with limited resources.
Abstract
BACKGROUND: Therapeutic drug monitoring (TDM) is essential for drugs with narrow therapeutic ranges and high pharmacokinetic variability. However, significant barriers hinder TDM implementation. Outsourcing of testing can delay timely clinical decision-making, with possible delays of up to 1 week. Additionally, management of analyte-specific certified reference material (CRM) presents substantial operational challenges, particularly for unstable compounds. The present study therefore developed and validated a high-performance liquid chromatography (HPLC) method based on relative molar sensitivity (RMS) of stable non-analyte reference materials (carbamazepine and caffeine) to quantify five clinically important analytes-acetaminophen, two beta-lactam antibiotics (cefmetazole and cefazolin), and amiodarone and its active metabolite, N-desethylamiodarone. METHODS: RMS values were calculated as the ratio of the analyte and non-analyte reference material calibration curve slopes. CRM solutions were used at 6 to 10 concentrations per analyte. HPLC analysis utilized analyte-specific detection wavelengths in the range 240-275 nm. For validation, control serum samples were spiked at 2 or 3 concentrations per analyte and processed via spin-column solid-phase extraction. Quantitative accuracy was assessed by comparing the RMS-based results with conventional absolute calibration methods. RESULTS: All calibration curves demonstrated excellent linearity, with correlation coefficients in the range 0.9993-1.0000. Calibration curve slope reproducibility was high, with relative standard deviations of 0.3%-1.5% (n = 3). Quantitative analyses of spiked serum samples showed close agreement between the RMS and conventional methods, with relative differences not exceeding ± 0.50% for either calibrant for any analyte. Both methods achieved comparable precision, with relative standard deviations of < 4.6%. CONCLUSIONS: The validated RMS-based HPLC method provides accurate and precise quantification of structurally diverse drugs using stable, the International System of Units (SI)-traceable, non-analyte reference materials. This approach eliminates the need for analyte-specific authentic standards for routine analyses, thus substantially reducing procurement costs and addressing the instability of some analytes. The rapid and accurate analysis enables timely reporting, potentially leading to improved clinical outcomes through expedited in-house TDM implementation in hospital laboratories with limited resources.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Evaluation and Improvement of the Analytical Method for Amiodarone Hydrochloride: Replacing the Pharmacopeial Method with a Validated RP-HPLC TechniqueApplied Sciences · 2026
- AQbD-Driven RP-HPLC Method for Rapid Dissolution Monitoring of Amiodarone Hydrochloride Tablets Without Sample Dilution.Pharmaceuticals (Basel, Switzerland) · 2026
- Volumetric Microsampling for Patient-Centric Therapeutic Drug Monitoring in Clinical Pharmacology: A Scoping Review.Journal of clinical pharmacology · 2026
- Stochastic evaluation of the reliability of relative molar sensitivity for quantitative HPLC analysis using a chemometric tool based on the FUMI theory.Journal of chromatography. A · 2026
- HPLC-MS Quantification of Multiple Tyrosine Kinase Inhibitors in Patients with Solid Tumors: Method Validation and Clinical Application.Pharmaceutics · 2026
- Development and validation of an HPLC-MS/MS method for the quantification of Aficamten in human plasma and urine and application to a clinical pharmacokinetic study.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026