Time-of-Day-Associated Variation in Serum CO2 Patient-Result Distribution and Its Impact on PBRTQC Performance.
Ziqing Jiang, Xuemei Wei, Lei Lin, Jie Feng, Yanhong Gao
Department of Clinical Laboratory, The First Medical Center of Chinese PLA General Hospital, Beijing 100853, China.
Key points
From the paper's conclusion- This study explores how time-of-day-associated variation in the serum CO2 patient-result distribution affects PBRTQC outcomes and clinical acidosis classification, and an optimization strategy is evaluated.
Abstract
Background: Patient-based real-time quality control (PBRTQC) is now widely used in clinical labs, yet the impact of time-dependent variation on its performance remains understudied, particularly with regard to serum carbon dioxide (CO2). This study explores how time-of-day-associated variation in the serum CO2 patient-result distribution affects PBRTQC outcomes and clinical acidosis classification, and an optimization strategy is evaluated. Methods: We analyzed 168,957 serum CO2 results collected over one year (2025) using three biochemical analyzers. Moving average-based PBRTQC was compared between peak (10:00-13:00) and off-peak periods, using either conventional unified limits or time-stratified separate limits. Results: Serum CO2 patient-result distribution shifted significantly toward higher values during the peak period (23.6 vs. 23.0 mmol/L, p < 0.001). A unified control strategy led to a marked rise in false-positive rates (0.955% vs. 0.062%). In a simulated category-transition analysis, 92 of 1212 samples (7.59%) were reclassified, of which 88 were acidosis-to-normal shifts. The time-stratified model reduced peak false positives to 0.069% while maintaining efficient detection of error. Conclusions: Time-of-day-associated variation in the distribution of patients' serum CO2 impairs PBRTQC specificity during peak hours. Time-stratified control limits reduce false-positive rates while maintaining error detection, offering a practical solution for reliable serum CO2 monitoring. These findings should be interpreted as descriptive, given the cross-sectional design and the potential influence of patient case mix.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.Diagnostics (Basel, Switzerland) · 2026
- Illustrative examples and data from patient-based real-time quality controlThe Clinical Biochemist Reviews · 2026
- B-239 Assessment of a novel PBRTQC approach based on Kullback-Leibler DivergenceClinical Chemistry · 2026
- B-043 Comparison of Turnaround Time of Beckman Coulter Power Processor and DxA 5000 Automated SystemsClinical Chemistry · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026