B-006 Polynomial Regression Analysis Techniques to Evaluate the Interchangeability of i-STAT® Cartridges and the Laboratory Method for Patient Blood Glucose Values. A Practical Example with Data Collect for Twelve Months

Valerio Genta, Brad Poore, Dina Radi, Sara Shumate, Carl Alston, David Mitchell

Key points

From the paper's conclusion
  • Conclusions The WPLS model showed that for a period of twelve months the relationship between paired patient blood glucose values obtained with i-STAT cartridges and the cobas 6000 method, was quasi-linear with few differences exceeding the criterion target value +/- 6mg/dl, greater.

Abstract

Background In emergency situations, i-STAT®1 cartridges offer rapid results at the patient bedside allowing both prompt medical decisions and interventions. In non-emergency situations, patient blood assays are performed with the Laboratory Method. To evaluate and monitor the interchangeability of i-STAT and the laboratory method for blood glucose values we compared the two methods with paired data obtained in one year using polynomial regression analysis techniques. Methods The patient blood glucose values obtained with i-STAT Chem8® or CG8® cartridges and those obtained with the laboratory method (LM) (Two cobas 6000®, Roche Diagnostics) within thirty minutes on the same patient, were collected electronically and transferred to Minitab® (Version 22, Minitab Inc) statistical software. Three thousand eight hundred seventy four paired observations obtained for twelve months were analyzed with orthogonal, ordinary (OLS), weighted polynomial (WPLS) least squares regression analysis, locally weighted scatter plot smoother (Lowess) statistical techniques, their diagnostics and graphic representations. The 99% confidence (CI) and 99% prediction (PI) intervals were obtained for the WPLS. The criterion target value +/- 6 mg/dL, or 8%, greater was used for acceptability of absolute and relative differences respectively. Results For blood glucose values between 30 and 600 mg/dL the orthogonal (i-STAT=1.3+1.006 LM) and the OLS (i-STAT=1.6+1.004 LM) regression equations were very similar. For both methods the standardized residuals by the fitted value showed increasing variance in Y for increasing values in X. Consequently, for additional analysis the WPLS was employed. WPLS equation: i-STAT=1+1.005 LM. The ANOVA table showed probabilistically significant differences between month regression lines (P<0.0001); this was due to very small Sy/x=0.4mg/dL. The lack-of-fit test showed probabilistically significant lack-of-fit (P<0.0001); this was due to very small pure error=0.1 mg/dL. The standardized deleted residuals (SDR) plots showed quasi-normal distribution, equality of variance and no unnatural distribution by observation order. The plot of SDR by the fitted value showed quasi-linear and similar lowess by month. The SDR plot by the leverage (Hi) showed a few possible outliers (SDR>|3|) and no influential observations (Hi<0.5). The CI and PI for laboratory method values of 50 mg/dL (Fit=52, CI=51-52,PI=45-58), 250 mg/dL (Fit=253, CI=252-253, PI=237-268) and 550 mg/dL (Fit=554, CI=552-555, PI-530-578) clearly indicated that the differences between methods would be within the total allowable error. This was corroborated by the plot of the relative differences by the month which showed that few relative differences exceeded the total error limits (69/3874=1.8%). Conclusions The WPLS model showed that for a period of twelve months the relationship between paired patient blood glucose values obtained with i-STAT cartridges and the cobas 6000 method, was quasi-linear with few differences exceeding the criterion target value +/- 6mg/dl, greater. Additionally, the 99% PI indicated that this relationship would be reliably anticipated if the same testing conditions were maintained. The automatic updating function of Minitab software allowed prompt detection of outliers, investigation and intervention by the Point of Care MLS. This study showed that the availability of multivariable statistical software, such as Minitab, was essential for the fast analysis of the data and their actionability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers