An improved approach to modelling patient reclassification: HbA 1c as an example
Christopher‐John L. Farrell, Tze Ping Loh, Tony Badrick, Graham R.D. Jones, Andrea R. Horvath
Royal North Shore Hospital · National University Hospital
Key points
From the paper's conclusion- We developed a method for this assessment based on reclassification of results at clinical decision limits and applied it to results from a haemoglobin A 1c (HbA 1c ) EQA program.
Abstract
Abstract Background External quality assurance (EQA) programs must assess whether the analytical performance of commonly used in vitro diagnostic (IVD) devices is fit for their intended clinical use. We developed a method for this assessment based on reclassification of results at clinical decision limits and applied it to results from a haemoglobin A 1c (HbA 1c ) EQA program. Content Using a commutable EQA program to provide analytical performance characteristics, we simulated 10,000 measurements of the HbA 1c value of each of 6,326 subjects in the National Health and Nutrition Survey (NHANES) by common IVD devices and the NHANES device. A parametric empirical Bayes approach was used to account for regression to the mean. We estimated the rates at which the IVD devices would reclassify results and compared them with the estimated reclassification rate for the NHANES device, calculating the difference between them as the ‘incremental reclassification rate’ (IRR). IRRs were ‘optimal’ if <1.5 %, ‘desirable’ if <3.0 %, and ‘minimal’ if <4.7 %. Summary and outlooks EQA performance data were available for five HbA 1c devices: Abbott Afinion, Bio-Rad D series, Roche Cobas, Sebia Capillarys, and Siemens DCA. The median IRRs were optimal for all methods for diabetes diagnosis (0.11–0.22 %) and for diabetes monitoring (−0.51–1.01 %), with negative IRRs representing lower reclassification rates than the NHANES device. For prediabetes diagnosis, the Afinion failed the minimal standard (8.62 %), but median IRRs were desirable for the other methods (1.43–2.69 %). EQA providers should consider reporting IRR for suitable analytes to help identify poorly performing IVD devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Translating analytical performance into clinical misclassification risk: a probabilistic framework for HbA1c interpretation.Clinical chemistry and laboratory medicine · 2026
- Impact of ISO/TS 20914-based measurement uncertainty on HbA1c diagnostic classification: A real-world comparison of two analytical systems.Clinica chimica acta; international journal of clinical chemistry · 2026
- Reliability maps for longitudinal monitoring: a probabilistic frameworkBiochimica Clinica · 2026
- Comparative analysis of enzymatic assay for HbA1c measurement with high performance liquid chromatography.Scientific reports · 2026
- Superiority of 1 h plasma glucose vs fasting plasma glucose, 2 h plasma glucose and HbA1c for the diagnosis of type 2 diabetesDiabetologia · 2025
- Diagnostic Utility of Fasting Blood Glucose and Glycated Hemoglobin Compared to Oral Glucose Tolerance Test for Diabetes Mellitus: A Cross-Sectional Study.Health science reports · 2026