Differences in the Glucose Management Indicator Between Continuous Glucose Monitoring Systems.
Katrine Bovién Gørlitz, Sara Heebøll, Morten Fenger-Grøn, Sanne Fisker, Søren Gregersen, Christian Trolle, Klavs Würgler Hansen
Medical Diagnostic Center, University Clinic for Innovative Patient Pathways, Regional Hospital Central Jutland, Silkeborg/Viborg, Denmark. · Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
Key points
From the paper's conclusion- Aim The glucose management indicator (GMI) estimates HbA1c from continuous glucose monitoring (CGM)‐derived mean glucose.
Abstract
AIM: The glucose management indicator (GMI) estimates HbA1c from continuous glucose monitoring (CGM)-derived mean glucose. We assessed whether this relationship differs between contemporary CGM systems. MATERIALS AND METHODS: In this observational study, mean sensor glucose was derived from 2 months of CGM data preceding HbA1c measurement in adults with Type 1 diabetes using Dexcom (Dx) G6/G7 (n = 252) or Medtronic Guardian 4/Simplera (GSi) (n = 180). Device-specific GMI equations were derived using linear regression of measured HbA1c on mean sensor glucose and compared with the original GMI equation. For direct comparison of the Dx and GSi GMI equations, these GMI equations were combined, with 95% confidence intervals calculated using the delta method. RESULTS: The Dx GMI equation yielded statistically significantly lower GMI values compared with the original GMI equation at mean sensor glucose levels < 13.7 mmol/L, with increasing differences at lower glucose levels. Conversely, at mean sensor glucose levels above 8.3 mmol/L, the GSi GMI equation yielded statistically significantly higher GMI values than the original GMI equation, with increasing differences at higher glucose levels. The Dx and GSi device-specific GMI equations yielded statistically significantly different GMI values > 6.7 mmol/L, with increasing differences at higher glucose levels. CONCLUSIONS: The relationship between mean sensor glucose and HbA1c differed between contemporary CGM systems. This finding suggests that device-specific GMI equations may improve interpretation of glycaemic control in the absence of standardisation across contemporary CGM systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Comparative Analysis Between Original and Updated Glucose Management Indicator and Glycated Haemoglobin Across Continuous Glucose Monitoring Sensors in Type 1 Diabetes.Diabetes, obesity & metabolism · 2026
- Updated glucose management indicator (GMI) better aligns with HbA1c than current GMI: implications for clinical practice and reportingDiabetologia · 2026
- Narrowing the A1c gap: Personalized modeling of HbA1c– continuous glucose monitor discordance in type 1 diabetesPLOS Digital Health · 2026
- High-Amplitude and Prolonged Glucose Excursions as a Key Determinant of Discordance Between Glucose Management Indicator and Glycated Hemoglobin in Type 1 DiabetesDiabetes Care · 2026
- A Multi-Site Prospective Review of the Use of Continuous Glucose Monitoring in People with HIV and Diabetes.AIDS patient care and STDs · 2026
- Association of HbA1c and an updated glucose management indicator (uGMI) with incident diabetic retinopathy in adults with type 1 diabetes: a longitudinal studyDiabetologia · 2025