A-002 Evaluation of the Abbott i-STAT devices for both high-sensitivity cardiac troponin I and creatinine
Abdullah Almeshari, Won-Shik Choi, Sameer Sharif, Shawn Mondoux, Alim Pardhan, Andrew Worster, Peter Kavsak
Key points
From the paper's conclusion- We retested this material on the same ARCHITECT analyzer in 2026 which yielded the following results: Low-QC=4.9 ng/L and High-QC=37.0 ng/L, indicating the cTnI epitope was stable and the material suitable for imprecision assessment.
Abstract
Background High-sensitivity cardiac troponin (hs-cTn) assays are now available to use in the near patient setting (e.g., at the bedside). These point of care (POC) tests are often performed on hand-held devices and hold promise for early decision making in emergent and urgent care settings. However, there are other tests that may be important and aid in risk stratification for patients with possible acute coronary syndrome. One such test is creatinine as the relationship between cardiac and kidney function is important when interpreting risk (Clin Chim Acta 2018;479:166-170). To that end, our objective was to evaluate a POC device that can perform both creatinine and hs-cTn testing in whole blood. Methods For this evaluation, we used the Abbott i-STAT 1 and i-STAT Alinity devices (hs-cTnI measuring range:2.9-1000ng/L; with values <2.9 set to 2.8). Imprecision was assessed over 25 days with quality control (QC) plasma pools (n=21 for both levels) for hs-cTnI (Clin Chim Acta 2019;498:27-29) and tricontrol levels for creatinine (n=20 Abbott i-STAT Level 1 Lot#301190 & Level 3 Lot#321188). The plasma pools were constructed in 2014 (High-QC) and 2015 (Low-QC) and stored frozen (below -700C) with the Abbott ARCHITECT hs-cTnI assay overall mean|SD|number of results being: Low-QC=4.1 ng/L|0.6|899 and High-QC=35.8 ng/L|2.6|804. We retested this material on the same ARCHITECT analyzer in 2026 which yielded the following results: Low-QC=4.9 ng/L and High-QC=37.0 ng/L, indicating the cTnI epitope was stable and the material suitable for imprecision assessment. In the United States only EDTA plasma is approved for testing with the Abbott ARCHITECT hs-cTnI assay whereas the i-STAT hs-cTnI assay is approved on whole blood and lithium heparin plasma. A method comparison study was performed on 120 EDTA plasma samples between the ARCHITECT and i-STAT (Passing-Bablok regression with 95% confidence intervals, CI) and between whole blood and lithium heparin plasma for i-STAT hs-cTnI and creatinine assays (n=20 samples tested on i-STAT 1 and Alinity) (note creatinine sample type is whole blood on i-STAT). The median differences (95%CI) were determined between whole blood and plasma with all analyses performed with Analyse-it software with acceptable hs-cTnI precision performance being a standard deviation =1.0 ng/L for the Low-QC and a coefficient of variation (CV) =10% for the High-QC (Clin Chem 2025;71:332-334). Results The imprecision for hs-cTnI on the i-STAT 1 (Low-QC mean|SD=4.3 ng/L|1.0 and High-QC mean |CV=38.6 ng/L|4.9%) and i-STAT Alinity (Low-QC mean|SD= 4.2 ng/L|0.8 and High-QC mean|CV =36.2 ng/L|4.7%) was acceptable as was creatinine (CVs ranged from 1.9-6.1%). Regression analyses indicated lower results with the i-STAT as compared to the ARCHITECT hs-cTnI assay in EDTA plasma (hs-cTnI range=1.4-136 ng/L) with slope=0.71 (95%CI:0.63-0.81) and intercept=0.7 (95%CI:0.6-0.8). The median difference (lithium heparin plasma from whole blood) was 1.1% (95%CI:-3.3 to 6.0) for hs-cTnI and 4.4% (95%CI:1.6to 6.2%) for creatinine on the i-STAT. Conclusion The i-STAT hs-cTnI imprecision, its comparability to ARCHITECT, and agreement between whole blood and plasma indicates suitable use and supports further testing whole blood with the i-STAT in a novel laboratory pathway (Clin Chim Acta 2026;579:120630) for risk stratification in the urgent/emergent setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world analytical and diagnostic evaluation of the Abbott i-STAT Alinity point-of-care high-sensitivity troponin I assay in a UK emergency department.Annals of clinical biochemistry · 2026
- A-375 Verifying analytical performance of i-STAT high sensitivity troponin cartridges in an urban academic medical center emergency departmentClinical Chemistry · 2026
- A-041 Clinical Performance and Risk Classification Agreement of Point-of-Care i-STAT High-Sensitivity Troponin I Compared with Abbott Alinity hs-cTnI in ED Patients with Suspected Acute Coronary SyndromeClinical Chemistry · 2026
- Validation of the eCL8000i point-of-care hs-cTnT assay: analytical performance and clinical utility in Chinese patientsJournal of Laboratory Medicine · 2026
- Diagnostic Performance of Initial High-Sensitivity Troponin I in Patients with Renal Dysfunction: A Systematic Review and Narrative Synthesis.Medicina (Kaunas, Lithuania) · 2026
- Interaction between creatinine and postoperative high-sensitivity troponin elevation for long-term mortality risk after non-cardiac surgery: a postoperative ICU cohortEuropean Heart Journal · 2026