A-360 Performance of rapid glucose and ketone screening methods in post-mortem vitreous fluid samples to determine ketoacidosis-mediated death
Jeslyn Chow, Rawan Alsafadi, Victoria Higgins, Matthew Nichols, Jennifer Dmetrichuk, Paul Yip
Key points
From the paper's conclusion- Overall, while POCT can support more timely forensic assessment, the variability observed with dipstick-based tests suggests they would be better suited as adjunctive tools and have limited impact on improving efficiency as standalone methods.
Abstract
Background Ketoacidosis (KA) is a potentially fatal condition resulting from excessive ketone production, leading to metabolic acidosis. Diabetic KA (DKA) is a specific form seen in diabetes, defined by elevated ketones and hyperglycemia. In forensic investigations, laboratory testing often takes weeks to months before results become available to inform the cause of death and ultimately be communicated to families of the deceased. Point-of-care testing (POCT) for ketones and glucose in post-mortem vitreous fluid (VF) could expedite forensic investigations through rapid screening tools such as glucometers and dipstick-based glucose and ketone tests. This study aims to compare the performance of these POCT methods to traditional laboratory methods along with their ability to identify KA during post-mortem assessment. Methods A total of 50 autopsy cases, including both KA (n=17) and non-ketoacidosis (n=33) confirmed deaths were assessed. Decedent VF glucose (n=48) was measured using the Contour Next glucometer (Ascensia Diabetes Care), while decedent VF glucose (n=22) and VF ketones (n=41) were assessed using the Keto-Diastix test strips detecting acetoacetate and acetone (Ascensia Diabetes Care). Laboratory methods were validated assays for glucose (Abbott Alinity hexokinase) and beta-hydroxybutyrate (BHB, Randox enzymatic on Abbott Architect). Receiver operator characteristic (ROC) curves were constructed to determine the ability of POCT to screen for DKA (n=15) and KA (all causes, n=17). Results ROC curve analysis revealed highest test efficiency for the glucometer VF glucose at a cutoff of >=21.9 mmol/L with a sensitivity of 87%, specificity of 100%, and area under the curve (AUC) of 0.962 to identify DKA. VF glucose dipstick testing demonstrated a sensitivity of 100% and specificity of 81% at a glucose cutoff of >=21.0 mmol/L, with an AUC of 1.00 for detection of DKA. Using these POCT cutoffs, overall qualitative agreement of glucometer VF glucose and dipstick VF glucose in comparison to the laboratory methods to screen for DKA was 100% and 86%, respectively, at a laboratory cutoff of >=20.0 mmol/L. Dipstick VF glucose testing resulted in 3 false positives relative to the laboratory method. VF ketone dipstick testing demonstrated a sensitivity of 83% and specificity of 90% at a cutoff of >=1.5 mmol/L, with an AUC of 0.908 for identifying KA (all causes). Using this cutoff, overall qualitative agreement was 73% relative to a laboratory threshold of >=3.0 mmol/L BHB to screen for KA. Dipstick VF ketone testing resulted in 4 false positives and 3 false negatives relative to the laboratory method. Conclusion Among the rapid screening tools evaluated, glucometer VF glucose demonstrated favorable screening performance for DKA-related death. However, some discordant results from dipstick testing suggest that dipstick-based methods remain susceptible to limitations including non-specificity and subjectivity of visual interpretation. Additionally, ketone dipsticks measure acetoacetate and acetone rather than BHB, the primary ketone in ketoacidosis, which may limit detection of lethal cases of KA before their appearance in VF. Overall, while POCT can support more timely forensic assessment, the variability observed with dipstick-based tests suggests they would be better suited as adjunctive tools and have limited impact on improving efficiency as standalone methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparing clinical laboratory beta-hydroxybutyrate testing of vitreous humor against toxicological blood analyses in ketoacidosis deaths.Journal of forensic sciences · 2026
- A-386 Evaluation of near patient beta hydroxybutyrate testing in post mortem samples using the StatStrip analyserClinical Chemistry · 2026
- B-406 The reliability of a significantly raised acetone (>10mg/100mL blood) to diagnose ketoacidosis in post mortem caseworkClinical Chemistry · 2026
- Detecting and subtyping ketoacidosis from metabolomic patterns in forensic caseworkScientific Reports · 2026
- B-028 Review of a sample exchange program for postmortem vitreous biochemical analysisClinical Chemistry · 2026
- A-159 Specificity of beta hydroxybutyrate in the confirmation of insulin-induced hypoglycemia in post mortem casesClinical Chemistry · 2026