A-372 Assessment of Urinary Albumin to Creatinine Ratio Using the Capitainer Dip70® Dried Urine Microsampling Device on Automated Clinical Analyzers

Mathias Karlsson, Johan Olausson, Dennis Poland, Tamara van Loosdregt, Anders Larsson

Key points

From the paper's conclusion
  • Keywords: albumin to creatinine ratio, dried urine microsampling, Dip70®, albuminuria, decentralized testing, kidney disease biomarkers

Abstract

Background Albuminuria is a central biomarker for early detection and monitoring of kidney damage, yet conventional liquid urine testing requires rapid processing, refrigeration, and cold chain logistics. These preanalytical dependencies limit large scale and decentralized screening. The Capitainer Dip70® is a novel dried urine microsampling device designed for fixed volume urine collection at ambient temperature. This study evaluated its analytical performance for measuring urinary albumin to creatinine ratio (ACR) on three major clinical chemistry platforms. Methods A total of 100 routine urine samples were collected as fresh liquid urine and matched Dip70® dried samples. After drying, the pre cut discs were eluted in PBS/Tween and analyzed on Roche Cobas Pro, Abbott Alinity, and Siemens Atellica chemistry analyzers. Albumin, creatinine, and ACR were compared between liquid and dried urine using linear regression, Bland–Altman analysis, and duplicate imprecision (CV%). Clinical agreement for albuminuria classification (normal, microalbuminuria, macroalbuminuria) was assessed using sensitivity and specificity. A stability study evaluated Dip70® performance over 30 days under four storage temperatures (-20°C, +4°C, room temperature, +37°C). Results ACR values from Dip70® samples showed strong analytical agreement with liquid urine across all platforms (Cobas Pro: R²=0.997; Alinity: R²=0.991; Atellica: R²=0.958). Regression slopes were close to unity, and biases were small (Bland–Altman bias: +0.41 to +0.60 mg/mmol). Duplicate precision for ACR was good, with total CVs of 3.7% (Cobas), 6% (Atellica), and 9.5% (Alinity). Classification accuracy for albuminuria was high. For mac¬ro¬albuminuria, sensitivity and specificity were 100% on cobas and Alinity, and 93%/100% on Atellica. Microalbuminuria sensitivity reached 100% for cobas and Atellica, while specificity ranged from 82% to 97%. In the stability study, ACR remained within the predefined acceptance limits (<15% deviation) for all temperatures up to 30 days except at +37°C, where deviations reached 15% at days 11 and 30. All trends showed modest negative drift over time, driven by temperature sensitive albumin degradation, while creatinine remained stable. Conclusions The Capitainer Dip70® provides a robust, user friendly alternative to conventional urine collection for ACR measurement. Across three clinical chemistry platforms, dried urine samples produced ACR values with excellent agreement, low imprecision, and reliable clinical classification. Storage stability under real world ambient conditions further supports its suitability for decentralized sampling, population screening, telemedicine, and remote patient monitoring. The fixed volume disc design minimizes subsampling errors common in dried spot methods and enhances reproducibility. Dip70® enables scalable, cold chain free albuminuria testing and has strong potential to expand access to kidney disease screening. Keywords: albumin to creatinine ratio, dried urine microsampling, Dip70®, albuminuria, decentralized testing, kidney disease biomarkers

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers