Quantification of Urinary NGAL Using Volumetric Dried Urine Sampling.
Susanne Sütterlin, Lena Carlsson, Ziad Amer, Sandra Näslund, Jesper von Groote, Anders Olof Larsson, Eva Tano
Department of Women's and Children's Health, Pediatric Nephrology and Infectious Diseases (PNID) Research Group, Uppsala University, 75185 Uppsala, Sweden. · Department of Medical Sciences, Section of Clinical Chemistry, Uppsala University, 75185 Uppsala, Sweden.
Key points
From the paper's conclusion- Conclusions: NGAL measurements obtained from volumetric dried urine sampling using Capitainer qDBS discs showed good agreement with liquid urine measurements.
Abstract
Objective: To evaluate the measurement of urinary neutrophil gelatinase-associated lipocalin (NGAL) from dried urine samples collected using Capitainer qDBS discs compared with liquid urine. Methods: Eighty urine samples from children aged 0-18 years were analysed. NGAL concentrations measured in liquid urine were compared with concentrations obtained from dried urine collected on Capitainer qDBS cards and eluted for analysis using a commercial ELISA. Method agreement was assessed using linear regression and Bland-Altman analysis. Results: Sixty-nine samples yielded measurable NGAL concentrations in both methods. Median NGAL concentrations were 8.72 ng/mL (IQR 3.97-20.88) in liquid urine and 8.48 ng/mL (IQR 3.28-20.11) in qDBS eluates. NGAL concentrations measured from qDBS correlated with liquid urine measurements with a strong linear relationship (R2 = 0.98). Method agreement was good. Median recovery of 95.4% indicated slightly lower concentrations in qDBS eluates. Conclusions: NGAL measurements obtained from volumetric dried urine sampling using Capitainer qDBS discs showed good agreement with liquid urine measurements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-372 Assessment of Urinary Albumin to Creatinine Ratio Using the Capitainer Dip70® Dried Urine Microsampling Device on Automated Clinical AnalyzersClinical Chemistry · 2026
- Dried Urine Spots for Quantification of C0 (Free) Carnitine Excretion: A Validated Method for Follow-Up and Confirmatory Testing of Primary Carnitine Deficiency.International journal of neonatal screening · 2026
- B-001 Limitations of Urine Dipstick Testing and the Clinical Value of Creatinine-Normalized Quantitative Urinalysis in Urologic and Renal Disease DetectionClinical Chemistry · 2026
- A-373 Creatinine Measurement Using the Dried Blood Microsampling Device Capitainer®B50: Analytical Validation, Correction Modeling, and Stability EvaluationClinical Chemistry · 2026
- Comparison of an automated and paper bladder diary in pediatric patients: mixed-methods, randomized crossover study.Frontiers in pediatrics · 2026
- Cross-Validation of Low-Cost Potentiometric and Colorimetric Methods for Urinary Urea Quantification: Toward Accessible Monitoring of Protein MetabolismAnalytica—A Journal of Analytical Chemistry and Chemical Analysis · 2026