A-368 An instrument-free, rapid finger-stick blood test for the quantitative determination of alanine aminotransferase (ALT), aspartate aminotransferase (AST), blood urea nitrogen (BUN) and creatinine (CRE)

Octavian Florescu, Ana Florescu, Allison Martin, Sunil Kurian, Alan Wu

Key points

From the paper's conclusion
  • Conclusion The Quic Liver and Kidney Panel has demonstrated strong clinical performance from a 50µL drop of whole blood.

Abstract

Background The American Liver Foundation estimates 100 million Americans suffer from liver disease, while the CDC estimates 36 million suffer from kidney disease. Yet a disproportionate number of those afflicted are elderly, rural and poorly mobile, resulting in delayed care, inferior outcomes and unnecessary costs. The Quic Liver and Kidney Panel is a rapid blood test for at-home use to facilitate the management of chronic liver and kidney disease. Plasma concentrations for ALT, AST, BUN and CRE are available from 50µL of whole blood obtained from a finger-stick or draw. Measurements are performed by an instrument-free, single-use disposable and results are transmitted through a low-cost USB-dongle to a phone or PC for display. Methods Within each disposable, whole blood is internally routed to multiple plasma separation membrane filters, where blood cells are trapped while the plasma flows passively by capillary action. The plasma is internally metered in 70nl cuvettes with 0.5mm path lengths +/-1%, loaded with dried reagents. Upon hydration of the dried reagents, coupled-enzyme reactions are initiated according to well established spectrophotometric reference methods. A microchip-integrated, 4-color, 8-cuvette spectrophotometer with narrowband LED illumination provides endpoint and rate measurements of the absorbances in each cuvette. UV photo-detectors and digital signal processing integrated on-chip calculate the analyte concentrations or activities from calibration values stored on-board. Quantitative results are displayed on the screen of a phone or PC with easy access to AI interpretation and tele-health support. Features like room temperature stability, self-heating, over- and under-fill control and auto-disabling of expired devices make the Quic system robust enough for at-home use. Results The Quic system demonstrated CVs of 11%, 6%, 9% and 4% (N=10), LOBs of 20 U/L, 22 U/L, 7 mg/dL and 0.2 mg/dL (N=10), LLODs of 32 U/L, 38 U/L, 8 mg/dL and 0.6 mg/dL (N=10) and ULODs above the specified targets of 1000 U/L, 1000 U/L, 115mg/dL and 12mg/dL (N=10), for ALT, BUN and CRE, respectively, on human clinical whole blood samples. The clinical performance of the Quic Liver and Kidney Panel was compared to the Abaxis Piccolo Xpress equivalent Abaxis Vetscan VS-2 on N=54 clinical human serum samples. The correlation coefficients (R2) for ALT, BUN and CRE were 0.99, 0.99, 0.95 and 0.96, respectively. The Quic system was subsequently compared to the Abaxis system on N=23 discarded same-day heparinized human whole blood samples at Zuckerberg San Francisco General Hospital. The correlation coefficients (R2) for ALT, BUN and CRE were 0.96, 1.00, 0.97 and 0.96, respectively. The study results are presented in the figure below. Conclusion The Quic Liver and Kidney Panel has demonstrated strong clinical performance from a 50µL drop of whole blood. The emphasis on microchip integration has eliminated the need for an instrument, making the Quic system ideal for at-home use. The ability to perform successive measurements mitigates sources of inter and intra biological variability, which can lead to earlier diagnoses, superior staging of chronic diseases and ultimately better outcomes. Funding for this project was provided in part through BARDA contract 75A50124C00013

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers