A-357 Characterization of Flea-Mixing-Induced Hemolysis in Blood-Gas Capillary Tubes
Lara Adib, Shankar Balasubramanian, Prasad Pamidi
Key points
From the paper's conclusion- This study evaluated the effect of flea-mixing on potassium (K+) and hemolysis results in capillary samples tested on the GEM Premier 7000 with iQM3 (GEM 7000).
Abstract
Background Metal “fleas” are commonly recommended for mixing capillary blood-gas samples to promote homogeneity and reduce microclot formation. Best practices for capillary flea-mixing techniques, particularly related to mixing frequency and intensity, are not well established due to the difficulty of assessing hemolysis in small-volume capillary tubes. This study evaluated the effect of flea-mixing on potassium (K+) and hemolysis results in capillary samples tested on the GEM Premier 7000 with iQM3 (GEM 7000). Methods Lithium heparinized venous blood was collected from five volunteers and transferred into 150-µL capillary tubes. Samples were subjected to controlled flea-mixing conditions: no flea-mixing (0 repetitions) and timed flea-mixing resulting in defined repetition counts. One repetition was defined as moving the flea from one end of the capillary to the other and back. Each condition was performed by at least two operators, and all samples tested in duplicate on the GEM 7000. Potassium concentrations and hemolysis values were extracted from the GEM7000 outputs. Within each donor-operator set, biases in K+ and hemolysis were calculated relative to the no-flea-mixing condition. Results A total of 146 samples were included in the analysis. Mixing up to 10 repetitions produced minimal changes in K+ and hemolysis compared with the no-flea-mixing condition. Mixing between 20 and 50 repetitions resulted in mean K+ and hemolysis biases of 0.18 mmol/L and 23.3 mg/dL, respectively. Prolonged mixing (51–130 repetitions) produced larger increases in both parameters, with mean K+ and hemolysis biases of 0.35 mmol/L and 61.8 mg/dL, respectively. Operator-to-operator variability in hemolysis was more pronounced at higher mixing frequencies. Conclusion Consistent with GEM 7000 recommendations, flea-mixing up to 10 repetitions does not meaningfully increase whole-blood potassium results or hemolyze samples and is appropriate for routine point-of-care practice to ensure sample homogeneity. Excessive flea-mixing is associated with elevated potassium and increased hemolysis results and should be avoided. Although venous blood was used in this evaluation, future studies will incorporate true capillary fingerstick samples to further validate these findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.European journal of pediatrics · 2026
- Whole blood hemolysis detection on the GEM® Premier™ 7000: performance across hemoglobin A, S, and F.Clinical chemistry and laboratory medicine · 2026
- A-409 Comparative analytical performance of cartridge-based blood gas analyzers and the clinical impact of integrated whole-blood hemolysis detectionClinical Chemistry · 2026
- When Hemolysis Misleads: Implications for the Clinical Interpretation of Enzyme Activities in Pediatric Samples.Medicina (Kaunas, Lithuania) · 2026
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026
- Phlebotomy technique and its impact on haemolysis in the emergency department.Annals of clinical biochemistry · 2026