A-409 Comparative analytical performance of cartridge-based blood gas analyzers and the clinical impact of integrated whole-blood hemolysis detection
Jacki Vollmer, Amy Wockenfus, Brandon Kelley, Brad Karon, Brooke Katzman
Key points
From the paper's conclusion- These findings indicate that BG and core laboratory results are not universally interchangeable, and platform-specific bias may influence clinical interpretation, particularly for Hb and K.
Abstract
Background Blood gas (BG) analyzers are essential point-of-care and central laboratory instruments used to assess acid–base status, ventilation, oxygenation, and electrolyte balance in critically ill patients. Cartridge-based systems are increasingly utilized due to simplified maintenance, integrated quality control, and rapid turnaround times. However, inter-instrument variability and method-specific bias remain important considerations, particularly when transitioning between platforms or comparing results with core laboratory analyzers. We evaluated two cartridge-based analyzers—Werfen GEM Premier 7000 (GEM 7000; Bedford, MA) and Siemens RAPIDPoint 500e (RAPIDPoint 500e; Malvern, PA)—compared with our current Radiometer ABL90 FLEX PLUS (ABL90; Brea, CA). Additionally, preanalytical factors—especially hemolysis—can significantly affect potassium and ionized calcium (iCa). Unlike central chemistry analyzers, most BG systems do not assess hemolysis in whole blood. GEM 7000 uniquely incorporates onboard semi-quantitative hemolysis detection. Methods Residual lithium-heparin whole blood syringes (N=120–130) were analyzed on all three platforms within 15 minutes. For analytes with established laboratory methods, paired specimens (N=31–33) were tested on Sysmex XN-1000 (hemoglobin [Hb]; whole blood) and Roche Cobas c501 (sodium [Na], potassium [K], glucose [Glu]; plasma). Acceptability was defined using CLIA Total Allowable Error (TEa) or internally established criteria where CLIA limits were unavailable. Hemolysis index (HI) comparisons (N=76) were performed between GEM 7000 (whole blood) and c501 (plasma following centrifugation). Selected samples were spiked with hemolysate to broaden the analytical range. Agreement was assessed using regression and categorical concordance. Results Compared with ABL90, GEM 7000 demonstrated =90% agreement within CLIA or internal TEa for pH, partial pressure of oxygen (pO2), partial pressure of carbon dioxide (pCO2), Hb, fractional carboxyhemoglobin (FCO2Hb), fractional methemoglobin (FMetHb), iCa, and Glu, but lower agreement for fractional oxyhemoglobin (FO2Hb), Na, K, and Lac. RAPIDPoint 500e achieved =90% agreement for pH, pO2, FMetHb, but showed reduced agreement for pCO2, FO2Hb, FCO2Hb, and Lac. (Table 1) Against primary laboratory methods, clinically relevant variability was observed. For Hb, percent within acceptance was 12.5% (ABL90), 62.5% (GEM 7000), and 28.1% (RAPIDPoint 500e). For Na, agreement was highest for ABL90 (96.8%) and lowest for GEM 7000 (64.5%). For K, GEM 7000 demonstrated strongest agreement (86.1%), whereas ABL90 showed the lowest (36.1%). HI correlation between GEM 7000 and c501 was strong (slope 1.005; mean difference 2), with 93.3% of samples within acceptance limits and 96.0% categorical agreement. Conclusion Both candidate analyzers demonstrated acceptable performance for core BG parameters; however, meaningful inter-platform variability was observed for co-oximetry, and Lac. These findings indicate that BG and core laboratory results are not universally interchangeable, and platform-specific bias may influence clinical interpretation, particularly for Hb and K. Institutions transitioning analyzers should consider method validation, clinician education, and potential adjustment of reference or critical value thresholds. The GEM 7000’s integrated hemolysis detection showed strong agreement with a laboratory-based plasma method, supporting its utility for identifying preanalytical interference at the point of testing. Real-time hemolysis assessment may reduce reporting of spurious results, limit unnecessary redraws, and enhance patient safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Whole blood hemolysis detection on the GEM® Premier™ 7000: performance across hemoglobin A, S, and F.Clinical chemistry and laboratory medicine · 2026
- Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.European journal of pediatrics · 2026
- A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage AnalyzerClinical Chemistry · 2026
- A-357 Characterization of Flea-Mixing-Induced Hemolysis in Blood-Gas Capillary TubesClinical Chemistry · 2026
- Pre-Analytical and Analytical Factors Affecting the Accuracy of Blood Gas Analysis—A Systematic Review with Structured Narrative SynthesisLaboratories · 2026
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026