A-286 Optimization of Ammonia Specimen Handling and Rejection Rates
Patrick Bokolo, Catherine Mills, Charles Keppler, Elizabeth Champion-Lyons, Nikolina Babic
Key points
From the paper's conclusion- We systematically evaluated ammonia stability across tube types and storage conditions, assessed the analytical and spectroscopic effects of hemoglobin, and examined how data driven testing and reporting strategies influence specimen rejection rates.
Abstract
Background Plasma ammonia measurement is critical for diagnosing and managing hyperammonemia in liver failure and metabolic disorders. Accurate results require rigorous control of preanalytical variables due to the inherent instability of this analyte. Clinical assay manufacturers advise against testing hemolyzed specimens because of threefold higher intracellular ammonia concentrations, yet potential spectrophotometric interference from hemoglobin remains incompletely characterized. Recent reports have identified lithium heparin specimens as suboptimal due to stability concerns. We systematically evaluated ammonia stability across tube types and storage conditions, assessed the analytical and spectroscopic effects of hemoglobin, and examined how data driven testing and reporting strategies influence specimen rejection rates. Methods Ammonia Stability Study: Blood was collected into lithium heparin and potassium EDTA tubes, placed on ice, centrifuged after 15 minutes, and evaluated for hemolysis index (HI) and ammonia at 2, 6, and 24 hours following refrigerated or frozen storage. Absolute changes from baseline were compared between tube types. Hemolysis Interference Study: Heparinized whole blood was washed, lysed, freeze-thawed, and centrifuged to generate hemolysate. Plasma-hemolysate mixtures spanning HI 80–290 were prepared and analyzed. Spectrophotometric effects of hemoglobin on ammonia assay performance were evaluated by extracting absorbance error data directly from the analyzer and stratifying error rates by HI. Rejection Rate Analysis: Ammonia and corresponding HI results from patient specimens (n = 4,278) were obtained from EPIC Workbench reports and analyzed. Rejection rates were evaluated across HI categories and stratified by age group for moderately hemolyzed specimens (HI 100–199). Results Ammonia Stability Study: Mean absolute change in ammonia from baseline across all storage times and conditions were 6.0 ± 2.8 µmol/L for lithium heparin and 6.8 ± 3.4 µmol/L for potassium EDTA tubes, indicating comparable stability. Hemolysis Interference Study: Hemolysis interference evaluation yielded measurable results up to HI 188, with mean ammonia of 206 ± 10 µmol/L and 3% bias relative to non-hemolyzed control, confirming no significant interference up to HI ∼200. All specimens with HI >188 generated absorbance errors, suggesting a spectrophotometric interference mechanism rather than intracellular ammonia at higher HI. Analysis of patient samples showed absorbance error rates <20% for HI <150, increasing to 55% for HI 150–200 and 88% for HI >200. Gross hemolysis spectrophotometric interference could be corrected by dilution. Rejection Rate Analysis: Overall rejection rate for all hemolyzed specimens was 24.0%. Restricting analysis to minimally hemolyzed samples (HI <100) reduced rejection rate to 7.0%, while inclusion of specimens with minimal to moderate hemolysis (HI 30–199) further lowered rejection rate to 2.0%. Conclusion This study demonstrates that ammonia can be reliably reported in mildly and moderately hemolyzed specimens (HI =200), with no significant contribution from intracellular ammonia. Implementation of this evidence based reporting threshold at our institution virtually eliminated ammonia specimen rejections by reducing rejection rate from 24% to 2%. Gross hemolysis may produce assay specific spectrophotometric interference, mitigated by dilution. Contrary to recent reports suggesting inferior stability of lithium heparin specimens, we observed comparable stability between lithium heparin and EDTA tubes, with no significant differences in temperature related handling effects. These findings support revised institutional practic
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-196 Irradiation-induced Ammonia in Sterilized EDTA Tubes: Multi-Tube Comparison and Evaluation of Irradiation-Related EDTA DegradationClinical Chemistry · 2026
- 12 Systematic Evaluation of Hemoglobin, Icterus, and Turbidity Interference in Urine Chemistry TestingAmerican Journal of Clinical Pathology · 2026
- A-213 Impact of Hemolysis on Serum Osmolality TestingClinical Chemistry · 2026
- Ammonia in Liver Disease: Diagnostic Utility, Preanalytical Challenges, Clinical Interpretation, and Therapeutic Implications: A Critical Narrative Review.Diagnostics (Basel, Switzerland) · 2026
- A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?Clinical Chemistry · 2026
- When Hemolysis Misleads: Implications for the Clinical Interpretation of Enzyme Activities in Pediatric Samples.Medicina (Kaunas, Lithuania) · 2026