A-196 Irradiation-induced Ammonia in Sterilized EDTA Tubes: Multi-Tube Comparison and Evaluation of Irradiation-Related EDTA Degradation

Sandra de Kozlowski, Regina Soldo, Snezana Jovicic, Kendall Cradic, Ana-Maria Simundic, Christian Moser

Key points

From the paper's conclusion
  • Across the 14 sterilized lots analyzed (from two manufacturers), K2EDTA and K3EDTA tubes exhibited ammonia concentrations ranging from approximately 2-12 µmol/L.

Abstract

Background Clinical ammonia testing is highly sensitive and requires strict preanalytical control. Ammonia concentrations can increase in whole blood due to delayed processing, inadequate cooling, and/or hemolysis. Tube types typically used for clinical ammonia testing are EDTA or lithium heparin. Blood collection tubes typically undergo sterilization by gamma, electron beam, or x-ray irradiation. Studies have shown that EDTA is highly unstable under gamma irradiation. This analytical, cross-sectional study analyzed multiple production lots of K2EDTA, K3EDTA, lithium heparin, sodium heparin, and no-additive tubes to characterize ammonia presence across tube types and evaluate the impact of irradiation. Methods An external certified laboratory (Linz, Austria) measured ammonia in June of 2018 using the Hach LCK 304 methodology. A total of 14 sterilized tube lots from two manufacturers, treated by either gamma or electron beam (E-Beam) irradiation, were analyzed. These included K2EDTA, K3EDTA and non-EDTA controls (no additive, sodium heparin). A separate internal experiment conducted in October 2018 at Greiner Bio-One R&D analyzed a single K2EDTA lot from a single manufacturer before and after electron-beam sterilization using the Hach LCK 304 methodology. All results were reported as µmol/L. Results The external laboratory’s results demonstrated that EDTA tubes contained measurable ammonium across multiple lots. Across the 14 sterilized lots analyzed (from two manufacturers), K2EDTA and K3EDTA tubes exhibited ammonia concentrations ranging from approximately 2-12 µmol/L. Non-EDTA tubes, including lithium heparin, and no-additive tubes, showed significantly lower concentrations at <1 to <3 µmol/L. The internal R&D results showed that ammonia concentrations increased from <1 µmol/L pre irradiation to approximately 6 µmol/L post irradiation. Conclusion This study demonstrates that EDTA blood collection tubes sterilized by irradiation contain measurable amounts of ammonia, irrespective of manufacturer. The pre/post irradiation experiment confirms that irradiation (Gamma and E-Beam) contributes to ammonia formation in EDTA tubes.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-196 Irradiation-induced Ammonia in Sterilized EDTA Tubes: Multi-Tube Comparison and Evaluation of Irradiation-Related EDTA Degradation — Science Skim