A-183 Standardizing Specimen Stability Studies for Venous Whole Blood, Serum, and Plasma: A Practical Framework with Analysis Methods, Acceptance Criteria, and Worked Examples
Marvin Berman, Patrick O’Donnell
Key points
From the paper's conclusion- By specifying key stability conditions, time/temperature durations, minimum time points, and choice of analysis method with predefined acceptance limits, the framework reduces ambiguity and accelerates evidence generation for specimen stability across venous whole blood, serum, and plasma
Abstract
Background Most clinical specimens undergo collection processing, transport, and storage prior to testing. Establishing specimen stability under these conditions is essential to avoid clinically meaningful bias. A standardized framework has been developed with support from regulatory and industry stakeholders, consolidating best practices for designing and analyzing specimen stability studies to support robust specimen labeling claims and regulatory submissions. Methods A cross disciplinary team synthesized current practices and CLSI aligned approaches to define: (1) study planning (intended population, specimen types, workflow mapping, risk assessment), (2) key factors affecting stability (temperature/relative humidity, time, light, freeze–thaw cycles, container interactions, sample handling), (3) experimental designs (classical vs. isochronous), and (4) two analysis strategies — linear regression (degradation modeling) and baseline comparison (mean shift or precent shift vs. baseline, two sided 95 percent CIs). A minimum of 4 time points including baseline, interim, claim, and claim plus delta time points are tested. Acceptance limits are set from total allowable error and clinical decision impact. Example conditions include room temperature (RT) bounds, 2–8?C, and frozen (less than 20?C); with a minimum of 3 samples spanning the lower, mid (near the medical decision limit) and upper region of the analytical measuring interval for a quantitative assay. For a qualitative assays, 3 samples should be evaluated below and above the medical decision cutoff. Results The framework provides operational templates for quantitative and qualitative assays, including sample size/replicate guidance, outlier handling, weighted vs. ordinary least-squares criteria, and claim reporting suitable for Instructions For Use (IFU) labeling. Worked examples illustrate stability designs for quantitative and qualitative measurands:Quantitative (Calcium): baseline comparison design across RT, 4?C, and frozen conditions; dating duration claims expressed per matrix (serum and plasma) within 10 percent acceptance linked to 20 percent total allowable error.Qualitative (HBsAg): on/off clot, RT and 4?C evaluations with percent agreement at negative/positive specimens bracketing the cutoff (C5/C95); dating duration claims expressed at RT and 4?C. An example risk assessment is provided (e.g., photosensitivity, evaporation, agitation/transport) with mitigations to preempt specimen instability. Conclusion This practical, adaptable framework enables developers to design scientifically rigorous, reliable specimen stability studies that translate directly into labeling claims and regulatory ready documentation. By specifying key stability conditions, time/temperature durations, minimum time points, and choice of analysis method with predefined acceptance limits, the framework reduces ambiguity and accelerates evidence generation for specimen stability across venous whole blood, serum, and plasma
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-322 Reducing test cancellations and redraws by extending manufacturer's stability claimsClinical Chemistry · 2026
- Pre-analytical stability of 7α-hydroxy-4-cholesten-3-one (7aC4): a systematic evaluation under clinically relevant storage conditions.Annals of clinical biochemistry · 2026
- 301 Stability of Routine Coagulation Assays at Room TemperatureAmerican Journal of Clinical Pathology · 2026
- Extended in-use stability of Krebs von den Lungen-6 (KL-6) reagents, quality controls, and calibrators under refrigerated conditionsInternational Medical Research Journal · 2026
- Effects of storage temperature and delayed analysis on routine biochemical and haematological test resultsScientific Reports · 2026
- Stability of 54 biochemical analytes in capped serum samples stored at 4-8 °C: a 72-h longitudinal evaluation.Clinical chemistry and laboratory medicine · 2026