Extended in-use stability of Krebs von den Lungen-6 (KL-6) reagents, quality controls, and calibrators under refrigerated conditions
Noor Hafizah Hassan, Ili Karmila Norazli, Siti Nurwani Ahmad Ridzuan, Muhammad Nursyazwan Zamre, Nurul Izzati Hamzan, Hamadah Mohammad Shariff, Fadzlyasraf Shaari, Hanna Nadhirah Mohd Radzi,
Institute for Medical Research · University of Selangor
Key points
From the paper's conclusion- This study evaluated the extended in-use stability of Nanopia KL-6 reagents, QCs, and calibrators for up to four months under refrigerated conditions.
Abstract
Reliable diagnostic testing depends on the stability of assay compo-nents, including reagents, quality controls (QCs), and calibrators. For the Nanopia KL-6 assay used in interstitial lung disease (ILD) assessment, the manufacturer states that in-use stability is limited to 1 month for reagents and QCs and approximately 2.5 months for calibrators at 2–8°C. This study evaluated the extended in-use stability of Nanopia KL-6 reagents, QCs, and calibrators for up to four months under refrigerated conditions. Extended stability testing was performed at seven prede-fined time points (T0–T6) over 132 days. Reagent stability was assessed using in-use QC materials at two concentration levels. QC stability was evaluated by aliquoting and testing stored QC material over time. Cali-brators stability was assessed across four concentration levels using bias analyses. Reagents and QCs demonstrated acceptable analytical stability for up to 132 days, with increasing bias observed beyond Day 90. All calibrator levels remained stable for up to 132 days with minimal bias (<10%). Nanopia KL-6 reagents, QCs and all calibrator levels remain sta-ble for up to 132 days when stored at 2–8°C, provided appropriate qual-ity monitoring is implemented.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-183 Standardizing Specimen Stability Studies for Venous Whole Blood, Serum, and Plasma: A Practical Framework with Analysis Methods, Acceptance Criteria, and Worked ExamplesClinical Chemistry · 2026
- Analytical validation of capillary blood Krebs von den Lungen-6 measurement using a high- sensitivity chemiluminescent immunoassay in pediatric patients.Frontiers in medicine · 2026
- Stability of 54 biochemical analytes in capped serum samples stored at 4-8 °C: a 72-h longitudinal evaluation.Clinical chemistry and laboratory medicine · 2026
- Pre-analytical stability of 7α-hydroxy-4-cholesten-3-one (7aC4): a systematic evaluation under clinically relevant storage conditions.Annals of clinical biochemistry · 2026
- Stability of serum calprotectin as an emerging inflammatory biomarker: a five-day study at -20 °C and -80 °C using two analytical methods.Laboratory medicine · 2026
- Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method.Diagnostics (Basel, Switzerland) · 2026