Clinical Chemistry

A-322 Reducing test cancellations and redraws by extending manufacturer's stability claims

Katherine Turner, Jennifer Elders

Key points

From the paper's conclusion
  • As a result of extending the stability we were able to decrease the amount of test cancellations and redraws.

Abstract

Background Adherence to pre-analytical specimen processing and stability requirements as defined by the manufacturer of FDA-approved tests is critical for accurate result reporting. The Immunodiagnostic Systems tests for human Growth Hormone (hGH), Insulin-Like Growth Factor-1 (IGF-1), Procollagen I Intact N-Terminal (PINP), and Beta-CrossLaps (CTX) require immediate separation of serum from the cells and must be frozen for transportation to the testing laboratory. Observance of this pre-analytical requirement can be challenging resulting in unacceptable specimens arriving at the laboratory, test cancellations, and patient redraws. Herein, we assess the room-temperature (RT) and refrigerated stability of separated serum for the above-mentioned tests to extend the manufacture’s stability claims and decrease the amount of test cancellations and redraws. Methods Low, medium, and high concentration pools were made for each analyte. Samples were tested in triplicate to obtain the initial concentration and then retested in triplicate after the following conditions: Eight hours at RT and 24-hours refrigerated. Acceptance criteria was less than a 10% difference between the initial concentration and the concentration obtained after storage. Results The percentage difference for all analytes tested was less than 10% for both RT and refrigerated storage conditions. The extended stability was applied only to specimens that did not maintain the primary stability requirement of frozen transport. Prior to extending the stability requirements, the mean (standard deviation) of test cancellations for unacceptable specimen over five months was 9.6 (7.0). Five months after the implementation, the mean (standard deviation) cancellation rate was 3 (2.1). Conclusion The manufacture of assays may provide limited stability claims, and it is up to the laboratory to define or extend those claims. With this project, we extended the stability claims to include RT and refrigerated transport temperatures of processed serum. As a result of extending the stability we were able to decrease the amount of test cancellations and redraws.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers