A-019 Analytical comparison of Roche cardiac troponin T (cTnT) on cobas e 801, e 602, and e 402
Brandon Kelley, Amy Wockenfus, Brad Karon, Leslie Donato
Key points
From the paper's conclusion- We compared cTnT results from Roche (Roche, Indianapolis, IN) cobas e 801 (801), e 601 (601), and two e 402 (402) instruments.
Abstract
Background Our institution has a cTnT biomarker panel consisting of an initial cTnT, followed by a 2-hour and if necessary, a 6-hour collection. Accurate cTnT results are important for rapid diagnosis of acute coronary syndrome and to avoid prolonged stays in an emergency room. We compared cTnT results from Roche (Roche, Indianapolis, IN) cobas e 801 (801), e 601 (601), and two e 402 (402) instruments. Methods Comparisons (N=91) were completed using cTnT results from a single 601 instrument as the reference and comparing to the results from two 402 analyzers. Additional comparisons (N=45) were completed using cTnT results from an 801 as reference compared to results from the 402 instruments. Residual plasma separator tubes (PST) from clinically ordered cTnT testing were used. To ensure sample quality all plasma samples were aliquoted and recentrifuged before testing on each analyzer. Analysis was completed on all analyzers within 1 hour of the first comparison measurement. Acceptance limits were set as Passing-Bablok slope 1.00 ±0.10 and bias ±5 ng/L (for cTnT values <100 ng/L) or ±5% (for cTnT values =100 ng/L) when compared to reference 601 or 801. Student’s T-Test was performed to assess statistical significance between the observed mean bias between Pure instruments compared to the reference instrument, significance was defined as p =0.05. Results When compared to 601, results from the first 402 instrument demonstrated a slope of 1.11 with 54% (49/91) of samples within our acceptance criteria and a mean difference of 115 ng/L (Figure 1). Results from the second 402 instrument had a slope of 1.11 with 53% (48/91) within acceptance limits with a mean difference of 121 ng/L (p = 0.53, no difference between 402 instruments). For samples with cTnT <100 ng/L, 88% (36/41) and 88% (36/41) were within acceptance. In cTnT samples =100 ng/L, the mean percent bias was observed as 12% on both 402 analyzers. When using results from the 801 instrument as reference, results from the first 402 instrument had a slope of 0.954 with 77% (33/43) within acceptable limits and a mean difference of -61 ng/L. Results from the second 402 instrument had a slope of 0.978 with 79% (34/43) within acceptable limits and a mean difference of -39 ng/L (p = 0.33, no difference between 402 instruments). In samples with cTnT <100, comparisons to the 801 showed 82% (14/17) and 94% (16/17) within acceptance criteria and a mean difference of 2 ng/L on each 402 instrument. For cTnT samples with concentrations =100 ng/L, the mean percent bias was -2% on both 402 instruments. Conclusion We observed proportional bias on the 402 instruments compared to both 601 and 801 instruments, with large and clinically significant differences particularly when comparing to the 601 at higher cTnT concentrations. To avoid a delta being caused by observed systematic bias we will complete all cTnT testing within a panel on one platform.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the eCL8000i point-of-care hs-cTnT assay: analytical performance and clinical utility in Chinese patientsJournal of Laboratory Medicine · 2026
- A-028 Real-world comparative clinical performance of high-sensitivity cardiac troponin T Gen5 and Gen6 in routine acute care testingClinical Chemistry · 2026
- Analytical performance of the Roche Elecsys TnT hs Gen 6 immunoassayClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 AssayThe Journal of Applied Laboratory Medicine · 2026
- A-033 Multisite, chronological (2015-2024) determinations of total hs-cTnT variation (analytical, biological, and preanalytical) demonstrate post-COVID increases in total intrapatient cTnT variation that degrade the optimal interpretation of low serial hs-cTnTClinical Chemistry · 2026
- A-013 Turnaround Time Performance for Troponin I in a Core LaboratoryClinical Chemistry · 2026