Validation of the eCL8000i point-of-care hs-cTnT assay: analytical performance and clinical utility in Chinese patients
Fei Ding, Yuyao Li, Jia Li, Minjie Zhang, Shanshan Liang, Yanhua Zhao
Key points
From the paper's conclusion- Conclusions The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage.
Abstract
Objectives We validated the analytical performance and clinical diagnostic efficacy of the eCL8000i point-of-care testing (POCT) system for high-sensitivity cardiac troponin T (hs-cTnT), following the standardized verification protocol for hs-cTnT POCT assays. Methods A two-phase evaluation was conducted: Phase I (564 samples) assessed precision, lower limit of quantitation (LoQ), method comparison, calibrator interchangeability, reference intervals and turnaround time (TAT) per CLSI guidelines; Phase II (721 emergency chest pain patients) used coronary angiography as the gold standard to evaluate AMI diagnostic performance vs. Roche Cobas e801 via receiver operating characteristic (ROC) analysis. Results Phase I showed total CVs of 5.22 % (low) and 1.69 % (high), qualified LoQ, strong correlation with Cobas e801 (r=0.998), acceptable calibrator interchangeability, sex-specific 99th percentile reference values, and TAT shortened to 17 min. Phase II revealed eCL8000i AUC 0.860 (95 % CI 0.833–0.885) vs. Cobas 0.891 (95 % CI 0.865–0.912); sensitivities/specificities were 74.8/85.1% and 82.0/84.3 %, with high negative predictive values (94.9 vs. 96.3 %). Conclusions The eCL8000i hs-cTnT POCT system meets international analytical standards and has reliable clinical performance for rapid AMI exclusion, complementing central laboratory testing for efficient chest pain center triage. Although the sample size for establishing sex-specific 99th percentile URLs (280 males, 284 females) did not fully meet the IFCC-recommended target of ≥400 per sex, our preliminary reference intervals provide a useful starting point for the Chinese population and require further validation in a larger cohort.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 AssayThe Journal of Applied Laboratory Medicine · 2026
- Clinical triage concordance of a point-of-care high-sensitivity cardiac troponin I assay using the ESC 0/1-hour algorithm.Clinica chimica acta; international journal of clinical chemistry · 2026
- A-375 Verifying analytical performance of i-STAT high sensitivity troponin cartridges in an urban academic medical center emergency departmentClinical Chemistry · 2026
- Utilization of Point-Of-Care Testing (POCT) or troponin in patients with acute chest painEuropean Heart Journal · 2026
- Diagnostic accuracy of high-sensitivity point-of-care troponin assays in emergency department for patients presenting with chest pain: a systematic review.European journal of emergency medicine : official journal of the European Society for Emergency Medicine · 2026
- A-041 Clinical Performance and Risk Classification Agreement of Point-of-Care i-STAT High-Sensitivity Troponin I Compared with Abbott Alinity hs-cTnI in ED Patients with Suspected Acute Coronary SyndromeClinical Chemistry · 2026