A-033 Multisite, chronological (2015-2024) determinations of total hs-cTnT variation (analytical, biological, and preanalytical) demonstrate post-COVID increases in total intrapatient cTnT variation that degrade the optimal interpretation of low serial hs-cTnT
Jenna Rimkus, Tar Aw, Yuelin Qiu, Jennifer Shea, Albert Tsui, Allison Venner, George Cembrowski
Key points
From the paper's conclusion- We recommend that laboratories determine their own total hs-cTnT variation from their last year of serial troponins under 20 ng/L and not from non-traceable quality control specim
Abstract
Background High-sensitivity cardiac troponin T (hs-cTnT) accelerates the diagnosis of suspected acute coronary syndrome, particularly with initial low troponin. During the COVID-19 pandemic, troponin elevations were more frequently observed with troponin T than troponin I and were associated with more frequent presumptive myocardial involvement. While contemporaneous publications demonstrated high hs-cTnT false positive rates, they were largely ignored. Last year, a Roche Canada bulletin described non-reproducible, falsely high troponins when hs-cTnT was used to analyze blood drawn into Lithium Heparin PST™ tubes. Roche identified White Particulate Matter—consisting of fibrinogen, platelets, and leukocytes—as a root cause. Using an algorithm that transforms serial intrapatient laboratory results into total variation [analytical + biological + preanalytical, (PMID:35137000)], we analyzed large, chronologically different sets of hs-cTnT measurements of blood drawn into PST™ or Barricor™ blood drawing tubes in Calgary, Saint John, Singapore and Edmonton hospitals. Methods Serial intrapatient hs-cTnT measurements (analyzed on the e801, e602, or Modular E170) were obtained from de-identified emergency room and intensive care unit patients. Calgary specimens were collected with PST™ blood drawing tubes, both pre-COVID (Dec 2017–July 2020) and post-COVID (Sep 2022–Apr 2024). Edmonton specimens (Barricor™ tubes) were collected between September 2022 and April 2024. Singapore data (PST™) were collected between July 2018 and July 2020. Saint John data included PST™ (May 2015–May 2018) and Barricor™ (May 2018–May 2021) collections. We studied hs-cTnT under 20 ng/L. Sequential intra-patient pairs of hs-cTnT were tabulated by between-test interval (hours). The Dahlberg calculation was used to determine the average within-pair variation for each interval. Linear regression of the variation curve to zero time eliminates the interference of diurnal variation, allowing the determination of total hs-cTnT variation: (s2analytical +s2preanalytical +s2biological )½ . Results Total hs-cTnT variation differed by time periods and blood drawing tube types. In Calgary (PSTTM), 23,666 pre-COVID patients yielded a variation of 9.2%, compared with 11,939 post-COVID patients with 18.5% (P<0.01). Edmonton (Barricor™) post-COVID demonstrated a variation of 6.3%, not exhibiting the same post-COVID increase. 8,270 Singapore patients (PST™, pre-COVID) demonstrated 11.1% total troponin variation. In Saint John, 1,811 PST™ patients yielded an 8.6% variation pre-COVID, whereas 1,970 Barricor™ patients demonstrated 5.9% variation during the pre-/mid-COVID period (P<0.01). The results are summarized in the Figure. Conclusion The post-COVID Calgary PST™ plasma troponin measurements demonstrate the highest total variation. This effect was not observed in Barricor™ plasma in Edmonton. Across sites, Barricor™ consistently yielded lower total variation than PST™, both pre- and post-COVID, with the difference becoming more pronounced in the post-COVID era. Given the clinical reliance on hs-cTnT, the increased post-COVID variability in measuring PST™ plasma will increase the risks of misclassifications in low troponin patients and associated adverse clinical decisions. These findings suggest post-COVID hs-cTnT interpretation using PST™ plasma may require reconsideration of post-COVID hs-cTnT variability expectations. We recommend that laboratories determine their own total hs-cTnT variation from their last year of serial troponins under 20 ng/L and not from non-traceable quality control specim
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-034 Post-COVID near-threshold instability with PST™ vs. Barricor™: implications for redundant testing and resource use.Clinical Chemistry · 2026
- Clinical Decisions and Outcomes After Switching High-Sensitivity Cardiac Troponin Assays in Suspected ACSJAMA Cardiology · 2025
- Long forms of cardiac troponin T for myocardial infarction diagnosis: the SuperTROPO studyEuropean Heart Journal · 2025
- A-023 Evaluation of Interfering Substances in VITROS High-Sensitivity Cardiac Troponin I AssayClinical Chemistry · 2026
- Clinical performance of a sixth-generation high-sensitivity cardiac troponin T assay compared with standard high-sensitivity troponin T in a real-world cohortEuropean Heart Journal · 2026
- Changes in high-sensitivity C-reactive protein testing and distribution across the COVID-19 pandemic: a 17-year laboratory-based study.Laboratory medicine · 2026