Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 Assay
Marko Knoll, Lori B. Daniels, Christian Mueller, Nicholas L Mills, Evangelos Giannitsis, Alisa F A Rösser, Dunja Kurtoic, Annika Wahl,
Roche (Switzerland) · University of California San Diego
Key points
From the paper's conclusion- Here, we characterize the analytical performance of a next-generation hs-cTn assay, Elecsys® Troponin T hs Gen 6 (Roche Diagnostics International).
Abstract
BACKGROUND: High-sensitivity cardiac troponin (hs-cTn) assays are recommended for the diagnosis of acute myocardial infarction. Here, we characterize the analytical performance of a next-generation hs-cTn assay, Elecsys® Troponin T hs Gen 6 (Roche Diagnostics International). METHODS: Surplus lithium-heparin plasma or serum samples from patients or healthy volunteers were run on Cobas® e 801, e 402, and Pro analyzers. Limits of blank (LoB), limits of detection (LoD), and limits of quantitation (LoQ) were determined according to CLSI EP17-A2, with target values of 1.0 and 1.5 ng/L for LoB/LoD and 3.0 ng/L (10% CV) and 1.5 ng/L (20% CV) for LoQ, respectively. Precision was measured, per CLSI EP17-A2, using 3 QC samples (approximately 4, 30, and 220 ng/L), 12 native samples, and 3 reagent lots. Linearity, per CLSI EP06-Ed2, was determined by diluting samples with cardiac troponin T (cTnT) concentration above the measuring range with a low/blank sample. Interference (per Glick) with endogenous and assay components at 5 cTnT concentrations was assessed. RESULTS: Measured values for LoB, LoD, and LoQ at 10% and 20% CV were 0.1 to 0.7 ng/L, 0.3 to 1.4 ng/L, 1.0 to 2.9 ng/L, and 0.4 to 1.2 ng/L, respectively. Repeatability CVs were 1.0 to 5.8% for mean cTnT concentrations of 2.6 to 9230 ng/L in lithium-heparin plasma. High precision was shown across lots, and linearity was observed across the measuring range (1.5 to 9500 ng/L, all Pearson's r = 1.00). No interferences were observed, specified up to ≤1000 mg/dL hemoglobin, ≤50 mg/dL [≤855 µmol/L] icterus/bilirubin, and ≤1200 ng/mL biotin. CONCLUSIONS: The analytical performance characterization of the assay demonstrated high sensitivity, high precision at the low end and across the measuring range, and resistance to interference.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Performance of the Sixth Generation High-Sensitivity Cardiac Troponin T AssayThe Journal of Applied Laboratory Medicine · 2026
- A-011 Analytical performance of the sixth generation high-sensitivity cardiac troponin T assayClinical Chemistry · 2026
- Analytical performance of the Roche Elecsys TnT hs Gen 6 immunoassayClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Clinical performance of a sixth-generation high-sensitivity cardiac troponin T assay compared with standard high-sensitivity troponin T in a real-world cohortEuropean Heart Journal · 2026
- Diagnostic validation of a novel high-sensitivity cardiac troponin T assayEuropean Heart Journal · 2026
- Establishing Reference Values in Healthy Participants for the Cardiac Troponin T High-Sensitivity Gen 6 Assay: REF-TSIX Global Reference StudyClinical Chemistry · 2026