Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 Assay

Marko Knoll, Lori B. Daniels, Christian Mueller, Nicholas L Mills, Evangelos Giannitsis, Alisa F A Rösser, Dunja Kurtoic, Annika Wahl, R. A. F. BODY, R H Christenson, Christa Cobbaert, Christopher R deFilippi, Kai M. Eggers, Kenji Inoue, A. S. Jaffe, C. F. McCarthy, James McCord, Johannes T Neumann, Torbjørn Omland, Cynthia Papendick, Yader Sandoval, Jack Wei Chieh Tan, Martin P Than, Raphael Twerenbold, W Frank Peacock, Steven J. R. Meex

原始摘要(英文原文)· Original abstract
BACKGROUND: High-sensitivity cardiac troponin (hs-cTn) assays are recommended for the diagnosis of acute myocardial infarction. Here, we characterize the analytical performance of a next-generation hs-cTn assay, Elecsys® Troponin T hs Gen 6 (Roche Diagnostics International). METHODS: Surplus lithium-heparin plasma or serum samples from patients or healthy volunteers were run on Cobas® e 801, e 402, and Pro analyzers. Limits of blank (LoB), limits of detection (LoD), and limits of quantitation (LoQ) were determined according to CLSI EP17-A2, with target values of 1.0 and 1.5 ng/L for LoB/LoD and 3.0 ng/L (10% CV) and 1.5 ng/L (20% CV) for LoQ, respectively. Precision was measured, per CLSI EP17-A2, using 3 QC samples (approximately 4, 30, and 220 ng/L), 12 native samples, and 3 reagent lots. Linearity, per CLSI EP06-Ed2, was determined by diluting samples with cardiac troponin T (cTnT) concentration above the measuring range with a low/blank sample. Interference (per Glick) with endogenous and assay components at 5 cTnT concentrations was assessed. RESULTS: Measured values for LoB, LoD, and LoQ at 10% and 20% CV were 0.1 to 0.7 ng/L, 0.3 to 1.4 ng/L, 1.0 to 2.9 ng/L, and 0.4 to 1.2 ng/L, respectively. Repeatability CVs were 1.0 to 5.8% for mean cTnT concentrations of 2.6 to 9230 ng/L in lithium-heparin plasma. High precision was shown across lots, and linearity was observed across the measuring range (1.5 to 9500 ng/L, all Pearson's r = 1.00). No interferences were observed, specified up to ≤1000 mg/dL hemoglobin, ≤50 mg/dL [≤855 µmol/L] icterus/bilirubin, and ≤1200 ng/mL biotin. CONCLUSIONS: The analytical performance characterization of the assay demonstrated high sensitivity, high precision at the low end and across the measuring range, and resistance to interference.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 Assay — 科研速览 Science Skim