The British journal of radiology

Comparison of Four CBCT-IGRT Geometric QA Phantoms within a Single Institutional Workflow.

Hyojun Park, Hyejo Ryu, Jin Hwa Choi, So-Yeon Park, Yoonsuk Huh, Seonghee Kang, Chang Heon Choi, Jung-In Kim,

Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do, 14353, Republic of Korea. · Department of Radiation Oncology, Chung-Ang University College of Medicine, Seoul, 06974, Republic of Korea.

Key points

From the paper's conclusion
  • The four phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks.

Abstract

OBJECTIVES: To compare four commercial cone-beam CT (CBCT) geometric quality assurance phantoms within a common workflow, focusing on positioning/repositioning accuracy and agreement between the laser-defined setup reference and the CBCT imaging isocentre. METHODS: MIMI™, QUASAR™ Penta-Guide, ISOCube, and MZ CUBE were evaluated monthly for five months on a Versa HD system with XVI CBCT and a HexaPOD couch. Two observers independently completed each measurement. Three-dimensional post-correction translational residuals (Δdpos/rep) and laser-defined setup reference-CBCT imaging isocentre distances (Δdiso) were calculated. Phantom comparisons used observer-averaged values from the five monthly sessions. Rotational correction and CBCT image quality were assessed descriptively with MZ CUBE. RESULTS: Mean Δdpos/rep values ranged from 0.19 to 0.31 mm, with mean absolute observer differences of 0.05-0.12 mm. No overall difference was found among the four phantoms (p = 0.547). Mean Δdiso values ranged from 0.68 to 0.87 mm, and all measurements were below 2 mm. An overall difference was observed for Δdiso (p = 0.019), although no pairwise comparisons were performed. With MZ CUBE, mean post-correction translational and rotational residuals were 0.24 mm and 0.37°, respectively, and image-quality measures showed no consistent temporal trend. CONCLUSIONS: The four phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks. ADVANCES IN KNOWLEDGE: Direct comparison within a common workflow showed similar translational performance, while the scope of additional testing depended on the phantom configuration and locally available accessories.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers