End-to-end technical validation of multi-isocenter total body irradiation using an anthropomorphic phantom.
Dipesh, Raj Pal Singh, Manindra Bhushan
Department of Physics, GLA University, Mathura, Uttar Pradesh, India. d.vashisht700@gmail.com. · Department of Physics, GLA University, Mathura, Uttar Pradesh, India.
Key points
From the paper's conclusion- The objective of this study is to perform an end-to-end (E2E) technical validation of a total body irradiation (TBI) workflow using an anthropomorphic phantom, integrating simulation, treatment planning, patient-specific quality assurance, and phantom-based dosimetry to ensure accurate and reproducible dose delivery.
Abstract
The objective of this study is to perform an end-to-end (E2E) technical validation of a total body irradiation (TBI) workflow using an anthropomorphic phantom, integrating simulation, treatment planning, patient-specific quality assurance, and phantom-based dosimetry to ensure accurate and reproducible dose delivery. A full E2E test was performed using the Rando Phantom and Varian Eclipse TPS. CT scans were acquired in head-first supine (HFS) and feet-first supine (FFS) orientations and registered to simulate full-body imaging. Treatment plans were optimized for both orientations, with specific gradient dose regions defined for overlap zones. ArcCHECK and OSLDs were used for pre-treatment and phantom-embedded dosimetric verification, respectively. Cone Beam CT (CBCT) assessed positional accuracy across eight treatment fractions. Dose discrepancies and alignment errors were evaluated across anatomical regions. HU comparison between the CT console and TPS revealed only minor deviations, all within ± 6 HU, supporting the consistency of HU representation in the TPS. Gamma passing rates exceeded 97% across most regions. OSLD dose differences remained within ± 2.5% for most sites, except thorax and overlap zones, which reached ± 4.8%. CBCT-based setup verification demonstrated sub-millimeter alignment deviations across fractions, indicating consistent positional accuracy. E2E validation using the Rando Phantom supports technical accuracy and consistency in the TBI workflow. Notable dose variations in overlap and heterogeneous regions highlight the need for enhanced QA in complex geometries. The findings support the technical feasibility and reproducibility of the workflow, while further patient-specific and motion-inclusive studies are required to establish broader clinical applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multicenter evaluation of total body irradiation planning strategies: A DGMP working group study on large-field irradiation techniques.Zeitschrift fur medizinische Physik · 2026
- Fictive Field-based Feathering for Extended-SSD Total Body Irradiation: A Single-institution Planning and Dosimetric Analysis.Journal of medical physics · 2026
- Quantifying Patient Setup Error in Total Body Irradiation With Megavoltage Computed Tomography-guided Helical Tomotherapy.In vivo (Athens, Greece) · 2026
- Total body irradiation practice patterns: insights from a national survey.Frontiers in oncology · 2026
- Feasibility, Treatment Planning, and Clinical Results of VMAT-Based Total Body Irradiation for Pediatric Patients: Single-Center Experience.Pediatric blood & cancer · 2026
- CT-ED/MD calibration curve commissioning and end-to-end validation of HyperSight CBCT-Based dose calculation for online adaptive radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026