B-346 Storage Stability of HbA1c in Whole Blood: Method-Dependent Interference on Ion-Exchange HPLC and Degradation Kinetics Characterized by LC–MS
Kuanysh Kabytaev, Zhuo Wang
Key points
From the paper's conclusion- We investigated HbA1c stability in intact versus freeze–thaw–lysed whole blood and characterized major hemoglobin adduct kinetics using LC–MS alongside routine ion-exchange HPLC assays.
Abstract
Background Preanalytical stability remains a major challenge in clinical laboratories, particularly in resource-limited environments. HbA1c is widely used in diabetes trials and is commonly measured in central laboratories using EDTA whole blood that is frozen and shipped on dry ice. Anecdotal observations suggest that freeze–thaw–induced hemolysis may reduce HbA1c stability and produce erroneous results on some ion-exchange HPLC systems, yet the underlying degradation processes and kinetics are not well described, especially using LC–MS. We investigated HbA1c stability in intact versus freeze–thaw–lysed whole blood and characterized major hemoglobin adduct kinetics using LC–MS alongside routine ion-exchange HPLC assays. Methods De-identified EDTA whole-blood specimens spanning a broad HbA1c range were frozen at -70 °C. After thawing, samples were kept at RT, and aliquots were withdrawn over intervals up to one week. In parallel, non-lysed whole-blood samples were stored under the same conditions and aliquoted the same way. HbA1c was measured by ion-exchange HPLC on the Tosoh G8 and Bio-Rad D-100. LC–MS was used to quantify a- and ß-chain modifications, including glycation, glutathionylation, and cysteinylation, enabling kinetic modeling of storage-related adduct formation. Results Freeze–thaw–lysed samples exhibited faster degradation and greater method-dependent HbA1c inaccuracy due to ion-exchange chromatographic interferences. On the Tosoh G8, the manufacturer-designated P00 peak (consistent with glutathionylated hemoglobin) increased over time and displayed saturating (“hockey-stick”) kinetics, while an intermittent peak to the right of the HbA1c window (P-HV3) caused unpredictable, clinically significant false HbA1c elevations (∼0.6 to >2.0% HbA1c units) independent of baseline HbA1c or storage duration. On the D-100, a degradation-associated peak (P3) accumulated similarly to the Tosoh P00 peak. LC–MS demonstrated a clearer overview of storage-related changes: glutathionylation increased over time, and no other major globin chain degradation products were observed. Interestingly, a- and ß-chain glycation ratios were stable across storage conditions, indicating that true glycation remained measurable by LC–MS even when ion-exchange HPLC results were compromised. In lysed samples, ß-chain glutathionylation increased over time and plateaued near ∼48 hours, with higher HbA1c specimens tending to show a lower maximal degree of glutathionylation. In intact whole blood, HbA1c was more stable, with negligible changes in globin adducts even after one week. Conclusion HbA1c stability is substantially reduced after freeze–thaw lysis compared with intact whole blood, and observed bias is strongly assay-dependent on ion-exchange HPLC. Progressive glutathionylation contributes to time-dependent bias, while sporadic co-eluting peaks (e.g., P-HV3) can generate large, unpredictable false HbA1c elevations in the Tosoh G8. LC–MS enables mechanistic characterization of degradation kinetics and can quantify original HbA1c due to the stability of a- and ß-chain glycated adducts even in highly degraded specimens.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of HbA1c stability in refrigerated blood samples: for how long are samples fit for use?Scandinavian journal of clinical and laboratory investigation · 2026
- Preanalytical stability of hemoglobin H inclusion body detection using rapid brilliant cresyl blue staining: a genotype-stratified stability study in EDTA blood stored at 4 °C.Laboratory medicine · 2026
- A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage AnalyzerClinical Chemistry · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?Clinical Chemistry · 2026
- 70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting IssuesAmerican Journal of Clinical Pathology · 2026