Design and statistical validation of low-cost electrogoniometers for elbow orthosis applications.
José F Escobedo-Magallán, Adriana Del Carmen Téllez-Anguiano, Manuel Áviles-Jr, Lorenzo Garcia
DEPI, TecNM, Instituto Tecnológico de Morelia, Avenida Tecnológico 1500, Col. Lomas de Santiaguito, Morelia, Mich., 58120, Mexico. · DEPI, Instituto Tecnológico de Morelia, Av. Tecnológico 1500, Col. Lomas de Santiaguito, Morelia, Michoacán, 58120, Mexico.
Key points
From the paper's conclusion- These results indicate that potentiometer-based electrogoniometers are suitable for wearable rehabilitation systems requiring reliable and realtime joint monitoring.
Abstract
This study presents the design, implementation, and statistical validation of four low-cost electromechanical electrogoniometers for elbow joint angle measurement in orthotic applications. The proposed systems are based on distinct sensing principles: an infrared (IR) optical sensor, a calibrated IR optical sensor, a rotary potentiometer, and a slide potentiometer, all integrated into a custom mechanical structure that enables indirect measurement while reducing mechanical stress on sensing components. Experimental validation was performed using a controlled elbow motion simulator and human subjects, with measurements compared against a reference digital goniometer. Performance metrics included mean absolute error (MAE), repeatability, linearity (R²), and statistical agreement. A one-way ANOVA revealed significant differences among sensing modalities (p < 0.05), and Tukey's post hoc analysis confirmed that the IR-based system exhibited significantly higher errors than calibrated IRbased and both potentiometer-based systems. Bland-Altman analysis demonstrated strong agreement for calibrated IR-based and potentiometer-based designs, with minimal bias and narrow limits of agreement. The calibrated IR achieved the highest accuracy (MAE ≈ 2.78°), while the slide potentiometer provided a favorable balance between performance (MAE ≈ 4.72°) and mechanical integration. These results indicate that potentiometer-based electrogoniometers are suitable for wearable rehabilitation systems requiring reliable and realtime joint monitoring. The proposed designs offer cost-effective alternatives for biomechanical instrumentation and support the development of accessible assistive technologies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wireless Instrumented Ankle Foot Orthosis (AFO) for Gait Cycle MonitoringInstruments · 2026
- A rigid-flexible coupled elbow rehabilitation monitoring system based on neural network fusionMeasurement Science and Technology · 2026
- A Brace-Integrated Fiber Bragg Grating Sensor with a Multi-Hardness Silicone Strain Transfer Substrate for Knee Flexion Monitoring: A Proof-of-Concept Study.Sensors (Basel, Switzerland) · 2026
- IMU- and Vision-Based Measurement Techniques for Joint Kinematics: A Narrative Review.Sensors (Basel, Switzerland) · 2026
- Sufficient Standardization? Evaluating the Reliability of an Inertial Sensor (BeyondTM) for Ankle Dorsiflexion After a Brief Familiarization PeriodSports · 2025
- A No-Code Toolchain for Validating Wearable Motion Sensing SystemsarXiv · 2026